Flat-Panel Volume Computed Tomography for Cochlear Implant Electrode Array Examination in Isolated Temporal Bone Specimens
Authors:
- Soenke H. BartlingSBSoenke H. Bartling
- Rajiv GuptaRGRajiv Gupta
- Attila TorkosATAttila Torkos
- Christian DullinCDChristian Dullin
- Grabbe EckhardtGEGrabbe Eckhardt
- Thomas LenarzTLThomas Lenarz
- Hartmut BeckerHBHartmut Becker
- Timo StöverTSTimo Stöver
- 1 June 2006
- journal article
- research article
- Published by Wolters Kluwer Health in Otology & Neurotology
- Vol. 27 (4) , 491-498
Abstract
Flat-panel based volume computed tomography could improve cochlear implant electrode evaluation in comparison with multislice computed tomography. Flat-panel based volume computed tomography offers higher spatial resolution and less metal artifacts than multislice computed tomography. Both characteristics could improve the evaluation of challenging but important questions in cochlear implantation assessment, such as an exact imaging of cochlea, osseous spiral lamina, electrode array position, and single electrode contacts. These questions are not currently fully answered by multislice computed tomography. Four isolated temporal bone specimens were scanned in a current multislice computed tomography scanner and in two experimental flat-panel based volume computed tomography scanners before and after cochlea implantation. To compare flat-panel based volume computed tomography and multislice computed tomography, four features were rated according to the following criteria: 1) visibility of the cochlea; 2) visibility of the osseous spiral lamina; 3) discernibility of individual electrode contacts; and 4) the ability to determine the electrode array position relative to scala tympani and scala vestibuli. Layer-by-layer microgrinding pictures were used as the ground truth for verification of imaging findings. Flat-panel based volume computed tomography was superior to multislice computed tomography in all four features rated. The cochlea and facial nerve canal were much better delineated in flat-panel based volume computed tomography. The osseous spiral lamina and single electrode contacts were only visible in flat-panel based volume computed tomography. Assessment of implant position with regard to the cochlear spaces was considerably improved by flat-panel based volume computed tomography. Cochlear implantation assessment could be improved by flat-panel based volume computed tomography and, therefore, would be highly beneficial for cochlea implantation research and for clinical evaluation. However, these first results were shown by scanning isolated temporal bone specimens; scanning whole human skull bases might be more challenging.Keywords
This publication has 14 references indexed in Scilit:
- Methodik der Evaluation perimodiolärer CI-Elektrodenträger im FelsenbeinmodellLaryngo-Rhino-Otologie, 2005
- Volume CT with a flat‐panel detector on a mobile, isocentric C‐arm: Pre‐clinical investigation in guidance of minimally invasive surgeryMedical Physics, 2005
- Quality Control after Cochlear Implant Surgery by Means of Rotational TomographyOtology & Neurotology, 2005
- Flat panel computed tomography of human ex vivo heart and bone specimens: initial experienceEuropean Radiology, 2004
- Ergebnisse mit dem Contour™-Cochlear-Implant bei KapselotoskleroseLaryngo-Rhino-Otologie, 2004
- Pediatric and Adult Cochlear ImplantationRadioGraphics, 2003
- Der Einsatz von Flachbilddetektoren f r die CT-BildgebungDer Radiologe, 2003
- CT-Derived Estimation of Cochlear Morphology and Electrode Array Position in Relation to Word Recognition in Nucleus-22 RecipientsJournal of the Association for Research in Otolaryngology, 2002
- Nondestructive Three-dimensional Analysis of Electrode to Modiolus ProximityOtology & Neurotology, 2002
- Cochlear view: Postoperative radiography for cochlear implantationAmerican Journal of Otolaryngology, 2000