Detective quantum efficiency of photon-counting CdTe and Si detectors for computed tomography: a simulation study
Open Access
- 17 July 2020
- journal article
- research article
- Published by SPIE-Intl Soc Optical Eng in Journal of Medical Imaging
- Vol. 7 (04) , 043501-28
Abstract
Photon-counting detectors are expected to become the next major advance in x-ray computed tomography (CT).1 – 3 Whereas the energy-integrating detectors in use today measure the total incident energy in the x-ray beam during each measurement time interval, photon-counting detectors are able to count the individual photons and measure their energy. This enables improved signal-to-noise ratio. Moreover, the energy-resolving capabilities can allow spectral imaging with better energy separation than what is practically achievable with existing dual-energy CT technologies, such as dual source,4 kVp-switching,5 split filtration,6 and dual-layer7 detectors. Photon-counting detectors can also provide higher spatial resolution.Keywords
All Related Versions
Funding Information
- National Institute of Biomedical Imaging and Bioengineering (U01 EB 017140)
This publication has 56 references indexed in Scilit:
- Accuracy of iodine quantification in dual-layer spectral CT: Influence of iterative reconstruction, patient habitus and tube parametersEuropean Journal of Radiology, 2018
- Initial Results of a Single-Source Dual-Energy Computed Tomography Technique Using a Split-FilterInvestigative Radiology, 2016
- Evaluation of conventional imaging performance in a research whole-body CT system with a photon-counting detector arrayPhysics in Medicine & Biology, 2016
- Vision 20/20: Single photon counting x‐ray detectors in medical imagingMedical Physics, 2013
- Toward quantifying the composition of soft tissues by spectral CT with Medipix3Medical Physics, 2012
- Quantification of head and body CTDIVOLof dual-energy x-ray CTwith fast-kVp switchingMedical Physics, 2011
- Energy-resolved computed tomography: first experimental resultsPhysics in Medicine & Biology, 2008
- Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomographyPhysics in Medicine & Biology, 2008
- Material differentiation by dual energy CT: initial experienceEuropean Radiology, 2006
- The influence of energy weighting on X-ray imaging qualityNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2004