Technical Mistakes during the Acquisition of the Electrocardiogram
- 4 October 2009
- journal article
- review article
- Published by Wiley in Annals of Noninvasive Electrocardiology
- Vol. 14 (4) , 389-403
Abstract
In addition to knowledge of normal and pathological patterns, the correct interpretation of electrocardiographic (ECG) recordings requires the use of acquisition procedures according to approved standards. Most manuals on standard electrocardiography devote little attention to inadequate ECG recordings. In this article, we present the most frequent ECG patterns resulting from errors in limb and precordial lead placement, artifacts in 12‐lead ECG as well as inadequate filter application; we also review alternative systems to the standard ECG, which may help minimize errors.Keywords
This publication has 65 references indexed in Scilit:
- Morfologías que indican colocación inadecuada de V1 y V2Revista Espanola de Cardiologia, 2008
- An accurate electrocardiographic algorithm for differentiation of tremor-induced pseudo-ventricular tachycardia and true ventricular tachycardiaInternational Journal of Cardiology, 2006
- Tremor as a Cause of Pseudo–Ventricular TachycardiaNew England Journal of Medicine, 1999
- Clinical Consequences of Electrocardiographic Artifact Mimicking Ventricular TachycardiaNew England Journal of Medicine, 1999
- RV6RV5 voltage ratio in systemic hypertensionThe American Journal of Cardiology, 1990
- Negative P wave in lead I due to right atrial enlargement and displacement of the heart in the chestLung, 1976
- Recommendations for Standardization of Leads and of Specifications for Instruments in Electrocardiography and VectorcardiographyCirculation, 1967
- The low-frequency response of electrocardiographs, a frequent source of recording errorsAmerican Heart Journal, 1966
- A new system of multiple-lead exercise electrocardiographyAmerican Heart Journal, 1966
- Standardization of precordial leadsAmerican Heart Journal, 1938