Validation of highly accelerated real‐time cardiac cine MRI with radial k‐space sampling and compressed sensing in patients at 1.5T and 3T
Authors:
- Jeremy D. CollinsJCJeremy D. Collins
- Edward V. R. DiBellaEDEdward V. R. DiBella
- Daniel C. LeeDLDaniel C. Lee
- Daniel KimDKDaniel Kim
- The CKD Optimal Management with Binders and NicotinamidE (COMBINE) Study GroupTGThe CKD Optimal Management with Binders and NicotinamidE (COMBINE) Study Group
- 17 September 2017
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 79 (5) , 2745-2751
Abstract
Purpose: To validate an optimal 12‐fold accelerated real‐time cine MRI pulse sequence with radial k‐space sampling and compressed sensing (CS) in patients at 1.5T and 3T.Methods: We used two strategies to reduce image artifacts arising from gradient delays and eddy currents in radial k‐space sampling with balanced steady‐state free precession readout. We validated this pulse sequence against a standard breath‐hold cine sequence in two patient cohorts: a myocardial infarction (n = 16) group at 1.5T and chronic kidney disease group (n = 18) at 3T. Two readers independently performed visual analysis of 68 cine sets in four categories (myocardial definition, temporal fidelity, artifact, noise) on a 5‐point Likert scale (1 = nondiagnostic, 2 = poor, 3 = adequate or moderate, 4 = good, 5 = excellent). Another reader calculated left ventricular (LV) functional parameters, including ejection fraction.Results: Compared with standard cine, real‐time cine produced nonsignificantly different visually assessed scores, except for the following categories: 1) temporal fidelity scores were significantly lower (P = 0.013) for real‐time cine at both field strengths, 2) artifacts scores were significantly higher (P = 0.013) for real‐time cine at both field strengths, and 3) noise scores were significantly (P = 0.013) higher for real‐time cine at 1.5T. Standard and real‐time cine pulse sequences produced LV functional parameters that were in good agreement (e.g., absolute mean difference in ejection fraction <4%).Conclusion: This study demonstrates that an optimal 12‐fold, accelerated, real‐time cine MRI pulse sequence using radial k‐space sampling and CS produces good to excellent visual scores and relatively accurate LV functional parameters in patients at 1.5T and 3T. Magn Reson Med 79:2745–2751, 2018. © 2017 International Society for Magnetic Resonance in Medicine.Keywords
Funding Information
- National Institutes of Health (R01HL116895, R01HL138578, R21EB024315, R21AG055954, R01DK102438)
This publication has 25 references indexed in Scilit:
- Improved temporal resolution in cardiac imaging using through‐time spiral GRAPPAMagnetic Resonance in Medicine, 2011
- Improved radial GRAPPA calibration for real‐time free‐breathing cardiac imagingMagnetic Resonance in Medicine, 2010
- Real‐time MRI at a resolution of 20 msNMR in Biomedicine, 2010
- ACCF/ACR/AHA/NASCI/SCMR 2010 Expert Consensus Document on Cardiovascular Magnetic ResonanceJACC, 2010
- High spatial and temporal resolution cardiac cine MRI from retrospective reconstruction of data acquired in real time using motion correction and resortingMagnetic Resonance in Medicine, 2009
- Fully automatic, retrospective enhancement of real‐time acquired cardiac cine MR images using image‐based navigators and respiratory motion‐corrected averagingMagnetic Resonance in Medicine, 2008
- Analysis and compensation of eddy currents in balanced SSFPMagnetic Resonance in Medicine, 2005
- Image quality assessment: from error visibility to structural similarityIEEE Transactions on Image Processing, 2004
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002
- Fast Fourier Transforms for Nonequispaced DataSIAM Journal on Scientific Computing, 1993