Development of an Integrated Disposable Device for SARSCoV-2 Nucleic Acid Extraction and Detection

Authors:

Abstract
Objective To develop a highly sensitive and rapid nucleic acid detection method for the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). Methods We designed, developed, and manufactured an integrated disposable device for SARS-CoV-2 nucleic acid extraction and detection. The precision of the liquid transfer and temperature control was tested. A comparison between our device and a commercial kit for SARS-Cov-2 nucleic acid extraction was performed using real-time fluorescence reverse transcription polymerase chain reaction (RT-PCR). The entire process, from SARS-CoV-2 nucleic acid extraction to amplification, was evaluated. Results The precision of the syringe transfer volume was 19.2 +/- 1.9 mu L (set value was 20), 32.2 +/- 1.6 (set value was 30), and 57.2 +/- 3.5 (set value was 60). Temperature control in the amplification tube was measured at 60.0 +/- 0.0 degrees C (set value was 60) and 95.1 +/- 0.2 degrees C (set value was 95) respectively. SARS-Cov2 nucleic acid extraction yield through the device was 7.10 x 106 copies/mL, while a commercial kit yielded 2.98 x 106 copies/mL. The mean time to complete the entire assay, from SARS-CoV-2 nucleic acid extraction to amplification detection, was 36 min and 45 s. The detection limit for SARS-CoV-2 nucleic acid was 250 copies/mL. Conclusion The integrated disposable devices may be used for SARS-CoV-2 Point-of-Care test (POCT).