An Implantable Electronic Device for Monitoring Fetal Lung Pressure in a Lamb Model of Congenital Diaphragmatic Hernia
- 18 January 2022
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 71 (00189456) , 1-10
Abstract
Tracheal occlusion (TO) is a fetal treatment option for congenital diaphragmatic hernia. It aims to reduce lung hypoplasia and hypertension by occluding the trachea with a subsequent increase in pulmonary pressure. However, how TO affects lung pressure is unknown, and no devices are available for its measurement. We developed an implantable device for real-time wireless monitoring and long-term logging of lung pressure in ovine fetuses. The device comprised a microcontroller unit, two pressure sensors, an inertial measurement unit (IMU), a radio frequency unit, and an antenna. In vitro and in vivo tests were performed to evaluate device performances. A diaphragmatic hernia was created in ovine fetuses via hysterotomy. Nine devices were implanted at 100 days of gestation and remained until delivery (~142 d). One pressure sensor measured pressure in the amniotic sac and the other in the fetal trachea. In vitro pressure accuracy was median (interquartile range) 0.31% (0.1%). In vitro device lifetime was ~60 days. The in vivo device recording time was 26.0 (3.9) days and it was well tolerated by the fetus and the ewe. Data telemetry operated throughout the device lifetime. At delivery, amniotic fluid, device encapsulation, and lamb skin did not present evidence of damages. Pressure data in four lambs were analyzed. Fast pressure changes due to the spontaneous fetal activity and long-term mean pressure changes resulting from TO were detected. The IMU allowed for the assessment of changes in fetal position. This device allows long-term monitoring of mean lung pressure, spontaneous breathing activity, and movements in physiological and pathological ovine fetal models.Keywords
Funding Information
- International Research Postgraduate Award from the University of Western Australia
- NHMRC Senior Research Fellowship (TNG1077691)
- Metropolitan Health Research Infrastructure Fund
This publication has 27 references indexed in Scilit:
- Temporary fetal tracheal occlusion using a gel plug in a rabbit model of congenital diaphragmatic herniaJournal of Pediatric Surgery, 2012
- Bio-magnetic signatures of fetal breathing movementPhysiological Measurement, 2011
- Technical aspects of fetal endoscopic tracheal occlusion for congenital diaphragmatic herniaJournal of Pediatric Surgery, 2011
- Tracheal occlusion: A review of obstructing fetal lungs to make them grow and matureAmerican Journal of Medical Genetics Part C: Seminars in Medical Genetics, 2007
- Towards a Miniature Implantable In Vivo Telemetry Monitoring System Dynamically Configurable as a Potentiostat or Galvanostat for Two- and Three-Electrode BiosensorsIEEE Transactions on Instrumentation and Measurement, 2005
- Survival of severe congenital diaphragmatic hernia has morbid consequencesJournal of Pediatric Surgery, 2005
- Totally implantable real-time in vivo video telemetry monitoring system for implant biocompatibility studiesIEEE Transactions on Instrumentation and Measurement, 2001
- A new photogrammetric method to measure fetal breathing movementsUltrasound in Obstetrics & Gynecology, 1999
- A transducer for detecting foetal breathing movements using PVDF filmPhysiological Measurement, 1993
- Characteristics of fetal lung fluid productionJournal of Pediatric Surgery, 1981