Two‐dimensional and serial column reversed‐phase separation of phenolic antioxidants on octadecyl‐, polyethyleneglycol‐, and pentafluorophenylpropyl‐silica columns
- 1 March 2006
- journal article
- research article
- Published by Wiley in Journal of Separation Science
- Vol. 29 (4) , 555-566
Abstract
The separation selectivity of octadecyl‐silica (C18) and of bonded pentafluorophenylpropyl‐silica (F5) and PEG‐silica columns was compared for natural phenolic antioxidants. The separation selectivities for phenolic antioxidants on C18 and F5 columns are strongly correlated, but low selectivity correlation indicating strong differences in the retention mechanism was observed between the C18 and PEG columns. Hence, the combination of a C18 and a PEG column is useful for separation of phenolic antioxidants that are not fully separated on single columns. Two‐dimensional comprehensive liquid chromatography using a short PEG‐silica column in the first dimension and a conventional C18‐silica in the second dimension has the advantage of on‐column focusing of the fractions transferred onto the C18 column in the second dimension, as a weaker mobile phase is used in the first dimension than in the second dimension. However, a stop‐flow set‐up in the first dimension system is necessary after the transfer of each fraction to the second dimension. Peak capacity is considerably larger but the separation time is much longer than with serially coupled PEG and C18 columns, which were employed for separation of beer and hop extract samples in connection with coulometric detection.Keywords
This publication has 59 references indexed in Scilit:
- The hydrophobic-subtraction model of reversed-phase column selectivityJournal of Chromatography A, 2004
- Phase System Selectivity and Peak Capacity in Liquid Column Chromatography ? the Impact on Two-Dimensional SeparationsChromatographia, 2004
- Determination of free and bound phenolic acids in beerFood Chemistry, 2004
- Characterization of the aromatic composition of some liquid foods by nuclear magnetic resonance spectrometry and liquid chromatography with nuclear magnetic resonance and mass spectrometric detectionAnalytica Chimica Acta, 2003
- Principal component analysis of the polyphenol content in young red winesFood Chemistry, 2002
- Analytical performance of commercially available and unavailable phenolic compounds using real samples by high-performance liquid chromatography–diode-array detectionAnalytica Chimica Acta, 2002
- Content of potentially anticarcinogenic flavonoids of 28 vegetables and 9 fruits commonly consumed in the NetherlandsJournal of Agricultural and Food Chemistry, 1992
- Comparison of electrochemical and ultraviolet detection methods in high-performance liquid chromatography for the determination of phenolic compounds commonly found in beers. Part 2. Analysis of beersThe Analyst, 1987
- Comparison of electrochemical and ultraviolet detection methods in high-performance liquid chromatography for the determination of phenolic compounds commonly found in beers. Part 1. Optimisation of operating parametersThe Analyst, 1987
- A new approach to the phenolic components in beer. Application to the determination of sinapic, ferulic, and p-coumaric acidsJournal of Agricultural and Food Chemistry, 1977