Abstract
Recently, new techniques for minimally invasive aortic valve implantation have been developed generating a need for planning tools that assess valve anatomy and guidance tools that support implantation under x-ray guidance. Extracting the aortic valve anatomy from CT images is essential for such tools and we present a model-based method for that purpose. In addition, we present a new method for the detection of the coronary ostia that exploits the model-based segmentation and show, how a number of clinical measurements such as diameters and the distances between aortic valve plane and coronary ostia can be derived that are important for procedure planning. Validation results are based on accurate reference annotations of 20 CT images from different patients and leave-one-out tests. They show that model adaptation can be done with a mean surface-to-surface error of 0.5 mm. For coronary ostia detection a success rate of 97.5 % is achieved. Depending on the measured quantity, the segmentation translates into a root-mean-square error between 0.4 − 1.2 mm when comparing clinical measurements derived from automatic segmentation and from reference annotations.Keywords
This publication has 9 references indexed in Scilit:
- Optimizing boundary detection via Simulated Search with applications to multi-modal heart segmentationMedical Image Analysis, 2010
- Multimodal Assessment of the Aortic Annulus DiameterJACC, 2010
- A planning system for transapical aortic valve implantation2009
- Localization and tracking of aortic valve prosthesis in 2D fluoroscopic image sequences2009
- Noninvasive Evaluation of the Aortic Root With Multislice Computed TomographyJACC: Cardiovascular Imaging, 2008
- Segmentation of the heart and major vascular structures in cardiovascular CT images2008
- Dynamic Model-Driven Quantitative and Visual Evaluation of the Aortic Valve from 4D CT2008
- Early experience with percutaneous transcatheter implantation of heart valve prosthesis for the treatment of end-stage inoperable patients with calcific aortic stenosisJACC, 2004