Clinical evaluation of the xDP output compression strategy for cochlear implants
Authors:
- 17 October 2015
- journal article
- research article
- Published by Springer Nature
- Vol. 273 (9) , 2363-2371
Abstract
Technological advances in the domain of digital signal processing adapted to cochlear implants (CI) are partially responsible for the ever-improving outcomes observed with this neural prosthesis. The goal of the present study was to evaluate audiometric outcomes with a new signal processing strategy implemented in Oticon Medical-Neurelec cochlear implant systems, the xDP strategy. The core of this approach is a preset-based back-end output compression system, modulating a multi-channel transfer function depending on the intensity and information content of input sounds. Twenty adult CI patients, matched for age and CI experience, were included in this study. Pure-tone thresholds and vocal audiometry scores were measured with their former signal processing strategy and with xDP. Speech perception was assessed using dissyllabic words presented in quiet, at different intensity levels: 40, 55, 70, and 85 dB SPL, and in a cocktail party noise at a signal-to-noise ratio of +10 dB. Results with the xDP strategy showed, as awaited, no major modification of pure-tone thresholds. A global increase of speech perception scores was observed after a 1-month habituation period, with significant improvements for speech perception in quiet for moderate (55 dB SPL), loud speech sounds (85 dB SPL), and speech-in-noise comprehension. Subjective signal quality assessment showed a preference for CrystalisxDP over the former strategy. These results allow the quantification of improvements provided by the xDP signal processing strategy.Keywords
Funding Information
- Oticon Medical-Neurelec
This publication has 29 references indexed in Scilit:
- Pre-, Per- and Postoperative Factors Affecting Performance of Postlinguistically Deaf Adults Using Cochlear Implants: A New Conceptual Model over TimePLOS ONE, 2012
- Comparison of dual-time-constant and fast-acting automatic gain control (AGC) systems in cochlear implantsInternational Journal of Audiology, 2009
- The Choice of Compression Speed in Hearing Aids: Theoretical and Practical Considerations and the Role of Individual DifferencesTrends in Amplification, 2008
- Mimicking the human earIEEE Signal Processing Magazine, 1998
- Speech recognition at simulated soft, conversational, and raised-to-loud vocal efforts by adults with cochlear implantsThe Journal of the Acoustical Society of America, 1997
- The Hearing Aid InputEar & Hearing, 1994
- Effect of compression ratio in a slow-acting compression hearing aid: Paired-comparison judgments of qualityThe Journal of the Acoustical Society of America, 1994
- Distribution of short-term rms levels in conversational speechThe Journal of the Acoustical Society of America, 1988
- A comparison of four methods of implementing automatic gain control (AGC) in hearing aidsBritish Journal of Audiology, 1988
- Statistical Measurements on Conversational SpeechThe Journal of the Acoustical Society of America, 1940