Comparison of Swept-Source Anterior Segment Ocular Coherence Tomography and Gonioscopy in Detecting Anterior Chamber Angle Closure
Open Access
- 1 February 2025
- journal article
- research article
- Published by Taylor & Francis in Clinical Ophthalmology
- Vol. ume 19, 699-711
Abstract
Purpose: To compare the diagnostic accuracy of swept-source anterior segment optical coherence tomography (AS-OCT) with gonioscopy in detecting angle closure among narrow-angle suspected eyes. Patients and Methods: A total of 125 eyes determined narrow anterior chamber angles suspected by Van Herick’s technique grade 0, 1, or 2 were recruited. AS-OCT was undergone before any contact procedure. Intraocular pressure (IOP) measurements using applanation tonometry and gonioscopy were performed, respectively. AS-OCT images were blindly interpreted by 3 glaucoma specialists, involving both qualitative assessments at 0, 90, 180, and 270 degrees and quantitative analysis using iris-trabecular contact (ITC) index. The sensitivity and specificity of AS-OCT, compared with gonioscopy—the gold standard for identifying anterior chamber angle closure—were calculated. Results: The mean age was 61± 12.9 years, and females constituted 67.6%. Eyes classified by Van Herick’s technique as grade 0, 1 and 2 were 16%, 58.40%, and 25.60%, respectively. Closed-angle identified by gonioscopy and AS-OCT were 100 eyes (80%) and 102 eyes (81.60%), respectively. The sensitivity and specificity of AS-OCT in detecting angle-closure were 92.16% (95% CI, 87.44– 96.87) and 73.91% (66.22– 81.61), respectively. For angle-closure identified by ITC index ≥ 55%, sensitivity was 86.36% (77.53– 95.20) and specificity was 92.86 (86.23– 99.49). The inter-observer agreement of gonioscopy was moderate (Kappa = 0.55), while the intra-observer and inter-observer agreement of AS-OCT was substantial (Kappa = 0.71– 0.80 and Kappa = 0.69, respectively). Conclusion: AS-OCT enables a contactless qualitative and quantitative assessment of angle-closure in narrow-angle suspected eyes, demonstrating high sensitivity, acceptable specificity, and good inter-observer and intra-observer reliability.Keywords
This publication has 21 references indexed in Scilit:
- Diagnostic accuracy of AS-OCT vs gonioscopy for detecting angle closure: a systematic review and meta-analysisGraefe's Archive for Clinical and Experimental Ophthalmology, 2021
- Association of Baseline Anterior Segment Parameters With the Development of Incident Gonioscopic Angle ClosureJAMA Ophthalmology, 2017
- Global Prevalence of Glaucoma and Projections of Glaucoma Burden through 2040Ophthalmology, 2014
- Assessment of Circumferential Angle-Closure by the Iris–Trabecular Contact Index with Swept-Source Optical Coherence TomographyOphthalmology, 2013
- Swept source optical coherence tomography measurement of the iris–trabecular contact (ITC) index: a new parameter for angle closureGraefe's Archive for Clinical and Experimental Ophthalmology, 2012
- Diagnostic Performance of Anterior Chamber Angle Measurements for Detecting Eyes With Narrow AnglesArchives of Ophthalmology (1950), 2010
- Screening for Narrow Angles in the Singapore Population: Evaluation of New Noncontact Screening MethodsOphthalmology, 2008
- Comparison of Gonioscopy and Anterior Segment Ocular Coherence Tomography in Detecting Angle Closure in Different Quadrants of the Anterior Chamber AngleOphthalmology, 2008
- Assessment of the Scleral Spur in Anterior Segment Optical Coherence Tomography ImagesArchives of Ophthalmology (1950), 2008
- Detection of Primary Angle Closure Using Anterior Segment Optical Coherence Tomography in Asian EyesOphthalmology, 2007