Superparamagnetic Iron Oxide-Enhanced Liver Magnetic Resonance Imaging

Authors:
  • JLJeong Min Lee
    Jeong Min Lee
  • MLMin Woo Lee
    Min Woo Lee
  • JCJin Young Choi
    Jin Young Choi
  • SKSe Hyung Kim
    Se Hyung Kim
  • JLJae Young Lee
    Jae Young Lee
  • JHJoon Koo Han
    Joon Koo Han
  • BCByung Ihn Choi
    Byung Ihn Choi

Abstract
We sought to compare the image quality, lesion conspicuity, and the diagnostic performance of 1.5 T and 3.0 T superparamagnetic iron oxide-enhanced liver magnetic resonance imaging (MRI) for detecting focal malignant hepatic lesions. A total of 35 patients with pathologically proven liver malignancy underwent both 1.5 and 3.0 T SPIO-enhanced MRI. The diagnostic accuracy was evaluated using the alternative-free response receiver operating characteristic method. Image artifacts, quality, and the lesion conspicuity were analyzed. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of the lesion were calculated. No significant difference of area under ROC curve (Az value) was noted. The mean SNR and CNR of the lesions was higher in the 3.0 T sets. There was no difference between the 1.5 T and the 3.0 T image sets for lesion conspicuity, but the image quality was better on 1.5 T. Motion and susceptibility artifacts were more frequent on 3.0 T. Diagnostic accuracies of the SPIO-enhanced MRI were equivalent on the 1.5 T and 3.0 T image sets. More prominent artifacts on 3.0 T superparamagnetic iron oxide-enhanced liver MRI counteracted advantage of higher SNR and CNR of 3.0 T.

This publication has 30 references indexed in Scilit: