The Elekta Fraxion™ system is not suitable for maxillary fixation in canine conformal radiation therapy techniques
- 18 January 2019
- journal article
- research article
- Published by Wiley in Veterinary Radiology & Ultrasound
- Vol. 60 (2) , 233-240
Abstract
In this prospective, exploratory study, we evaluated the positioning accuracy in a group of 15 dogs undergoing fractionated stereotactic radiotherapy for tumors affecting the head, using a modified human maxillary fixation device (Elekta Fraxion™ system). Positioning was assessed using on‐board volumetric imaging, with a six‐degrees‐of‐freedom image registration technique. Prior to treatment delivery, CBCT images were obtained and patient alignment was corrected, in both translational and rotational planes, using a six‐degrees‐of‐freedom robotic patient positioning system (HexaPOD Evo RT System). The maximum angular inter‐fraction motions observed were 6.1° (yaw), 10.9° (pitch), and 4.5° (roll). The mean systematic translational errors were 4.7, 2.6, and 2.3 mm, mean random translational errors were 3.0, 2.2, and 2.5 mm, and mean overall translational errors were 2.4, 0.7, and 2.3 mm in the cranial‐caudal, lateral, and dorsal‐ventral directions, respectively. The mean systematic rotational errors were 1.17°, 0.77°, and 1.43°, the mean rotational random errors were 1.65°, 1.46°, and 1.34° and the mean overall rotational errors were 0.56°, 0.22°, and 0.29° in the yaw, pitch, and roll directions, respectively. The mean error of the three‐dimensional vector was 6.9 mm with a standard deviation of 3.8 mm. Ninety‐five percent of the three‐dimensional vectors were <14.8 mm. This study demonstrates that this maxillary fixation device relies on six‐degrees‐of‐freedom registration and an ability to apply corrections using a six‐degrees‐of‐freedom couch for accurate patient positioning and tumor targeting. Its use in conformal radiation therapy in dogs is not recommended.Keywords
This publication has 20 references indexed in Scilit:
- A NOVEL SUPPORT DEVICE FOR HEAD IMMOBILIZATION DURING RADIATION THERAPY THAT IS APPLICABLE TO BOTH CATS AND DOGSVeterinary Radiology & Ultrasound, 2015
- VALIDATION OF AN INDEXED RADIOTHERAPY HEAD POSITIONING DEVICE FOR USE IN DOGS AND CATSVeterinary Radiology & Ultrasound, 2015
- Stereotactic radiation therapy for treatment of canine intracranial meningiomasVeterinary and Comparative Oncology, 2014
- Frameless stereotactic radiosurgery for the treatment of primary intracranial tumours in dogsVeterinary and Comparative Oncology, 2013
- Three‐dimensional conformal radiation therapy alone or in combination with surgery for treatment of canine intracranial meningiomasVeterinary and Comparative Oncology, 2013
- 4D registration and 4D verification of lung tumor position for stereotactic volumetric modulated arc therapy using respiratory-correlated cone-beam CTJournal of Radiation Research, 2012
- HELICAL TOMOTHERAPY SETUP VARIATIONS IN CANINE NASAL TUMOR PATIENTS IMMOBILIZED WITH A BITE BLOCKVeterinary Radiology & Ultrasound, 2012
- Assessment of three-dimensional set-up errors in conventional head and neck radiotherapy using electronic portal imaging deviceRadiation Oncology, 2007
- Errors and margins in radiotherapySeminars in Radiation Oncology, 2004
- Radiosurgery using a stereotactic headframe system for irradiation of brain tumors in dogsJournal of the American Veterinary Medical Association, 2001