The characterization and transmission of scattered radiation resulting from x-ray beams filtered with zero to 0.99 mm copper
Authors:
- D G SuttonDSD G Sutton
- C J MartinCMC J Martin
- D PeetDPD Peet
- J R WilliamsJWJ R Williams
- 3 May 2012
- journal article
- Published by IOP Publishing in Journal of Radiological Protection
- Vol. 32 (2) , 117-129
Abstract
The specification of shielding for fluoroscopic facilities in the UK is based on the determination of scatter incident on a barrier using a simple formula linking kerma area product and scatter kerma. Over the last few years there has been a move to incorporate additional copper filtration in equipment used for high dose fluoroscopic and interventional examinations, and the existing formula does not take this into account. The spectral and transmission characteristics of the scattered radiation resulting from a primary x-ray beam filtered with additional copper are not known. In this study, the relationship between primary and scattered radiation in these beams has been investigated as have their transmission characteristics. The scatter kerma area product from filtered beams is shown to be greater than that from conventional x-rays and a simple numerical correction linking the two has been derived. The implications for shielding calculations have been assessed and the impact of the correction on calculated barrier thickness has been shown to be relatively small. The broad beam transmission characteristics of the radiation scattered from the filtered beams have been simulated using Monte Carlo methods and found to be adequately described by the standard transmission equation using conventional coefficients.Keywords
This publication has 11 references indexed in Scilit:
- INVESTIGATION OF THE PRACTICAL ASPECTS OF AN ADDITIONAL 0.1 MM COPPER X-RAY SPECTRAL FILTER FOR CINE ACQUISITION MODE IMAGING IN A CLINICAL CARE SETTINGHealth Physics, 2010
- TECHNICAL ADVANCES OF INTERVENTIONAL FLUOROSCOPY AND FLAT PANEL IMAGE RECEPTORHealth Physics, 2008
- A large-scale multicentre study in Belgium of dose area product values and effective doses in interventional cardiology using contemporary X-ray equipmentRadiation Protection Dosimetry, 2007
- The operation logic of automatic dose control of fluoroscopy system in conjunction with spectral shaping filtersMedical Physics, 2007
- Optimization of radiographic parameters for paediatric cardiac angiographyBritish Journal of Radiology, 2004
- Fluoroscopy: Recording of Fluoroscopic Images and Automatic Exposure ControlRadioGraphics, 2001
- Secondary Shielding Barriers for Diagnostic X-Ray FacilitiesHealth Physics, 1998
- Evaluation of NCRP Report No. 49 assumptions on workloads and use factors in diagnostic radiology facilitiesMedical Physics, 1996
- Transmission Data for Shielding Diagnostic X-Ray FacilitiesHealth Physics, 1995
- Diagnostic X-ray Shielding Design Based on an Empirical Model of Photon AttenuationHealth Physics, 1983