Complications following intracranial pressure monitoring in children: a 6-year single-center experience
Authors:
- 1 March 2018
- journal article
- research article
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of neurosurgery. Pediatrics
- Vol. 21 (3) , 278-283
Abstract
OBJECTIVE: Intracranial pressure (ICP) monitoring is an important tool in the neurosurgeon’s armamentarium and is used for a wide range of indications. There are many different ICP monitors available, of which fiber-optic intraparenchymal devices are very popular. Here, the authors document their experience performing ICP monitoring from 2005 to 2015 and specifically complication rates following insertion of the Microsensor ICP monitor.METHODS: A retrospective case series review of all patients who underwent ICP monitoring over a 10-year period from 2005 to 2015 was performed.RESULTS: There were 385 separate operations with an overall complication rate of 8.3% (32 of 385 cases). Hardware failure occurred in 4.2% of cases, the CSF leakage rate was 3.6%, the postoperative hemorrhage rate was 0.5%, and there was 1 case of infection (0.3% of cases). Only patients with hardware problems required further surgery as a result of their complications, and no patient had any permanent morbidity or mortality from the procedure. Younger patients (p = 0.001) and patients with pathologically high ICP (13% of patients with high ICP vs 6.5% of patients with normal ICP; p = 0.04) were significantly more likely to have complications. There was no significant difference in the complication rates between general neurosurgical patients and craniofacial patients (7.6% vs 8.8%, respectively; p = 0.67).CONCLUSIONS: Intraparenchymal ICP monitoring is a safe procedure associated with low complications and morbidity in the pediatric craniofacial and neurosurgical population and should be offered to appropriate patients to assess ICP with the reassurance of the safety record reported in this study.Keywords
This publication has 17 references indexed in Scilit:
- The preoperative incidence of raised intracranial pressure in nonsyndromic sagittal craniosynostosis is underestimated in the literatureJournal of neurosurgery. Pediatrics, 2014
- An evaluation and comparison of intraventricular, intraparenchymal, and fluid-coupled techniques for intracranial pressure monitoring in patients with severe traumatic brain injuryJournal of Clinical Monitoring and Computing, 2011
- Diagnostic Intracranial Pressure Monitoring and Surgical Management in Idiopathic Normal Pressure HydrocephalusNeurosurgery, 2010
- External ventricular drains in pediatric patients*Pediatric Critical Care Medicine, 2009
- Risk factors and complications of intracranial pressure monitoring with a fiberoptic deviceJournal of Clinical Neuroscience, 2009
- ASSESSMENT OF ZERO DRIFT IN THE CODMAN INTRACRANIAL PRESSURE MONITORNeurosurgery, 2009
- External ventricular drain infection: the effect of a strict protocol on infection rates and a review of the literatureBritish Journal Of Neurosurgery, 2006
- Risk factors of intracranial pressure monitoring in children with fiberoptic devices: A critical reviewSurgical Neurology, 1997
- Continuous-Pressure Controlled, External Ventricular Drainage for Treatment of Acute Hydrocephalus—Evaluation of Risk FactorsNeurosurgery, 1992
- A comparison of intraparenchymatous and intraventricular pressure recording in clinical practiceJournal of Neurosurgery, 1987