Metallic Artifact Reduction in Midfacial CT Scans Using Patient-Specific Polymer Implants Enhances Image Quality
Authors:
Open Access
- 28 January 2023
- journal article
- research article
- Published by MDPI AG in Journal of Personalized Medicine
- Vol. 13 (2) , 236
Abstract
Midfacial reconstruction after tumor resection surgery is commonly conducted by using autologous bone grafts or alloplastic implants. Titanium is the most frequently used osteosynthesis material in these cases but causes disturbing metallic artifacts in CT imaging. The purpose of this experimental study was to evaluate whether the use of midfacial polymer implants reduces metallic artifacts in CT imaging to improve image quality. Zygomatic titanium (n = 1) and polymer (n = 12) implants were successively implanted in a human skull specimen. Implants were analyzed for their effect on Hounsfield Unit values (streak artifacts) and virtual growth in CT images (blooming artifacts) as well as image quality. Multi-factorial ANOVA and Bonferroni’s post hoc test were used. Titanium (173.7 HU; SD ± 5.1) and hydroxyapatite containing polymers (155.3 HU; SD ± 5.9) were associated with significantly more streak artifacts compared to all other polymer materials. There was no significant difference in blooming artifacts between materials. The metallic artifact reduction algorithm showed no significant difference. Image quality was slightly better for polymer implants compared to titanium. Personalized polymer implants for midfacial reconstruction significantly reduce metallic artifacts in CT imaging which improves image quality. Hence, postoperative radiation therapy planning and radiological tumor aftercare around the implants are facilitated.Keywords
Funding Information
- Federal Ministry of Education and Research (BMBF) (13XP5108A)
This publication has 43 references indexed in Scilit:
- Current Strategies to Improve the Bioactivity of PEEKInternational Journal of Molecular Sciences, 2014
- A matched‐pair comparison of two different locking plates for valgus‐producing medial open‐wedge high tibial osteotomy: peek–carbon composite plate versus titanium plateKnee Surgery, Sports Traumatology, Arthroscopy, 2014
- Experimental verification of ion stopping power prediction from dual energy CT data in tissue surrogatesPhysics in Medicine & Biology, 2013
- The impact of low-Z and high-Z metal implants in IMRT: A Monte Carlo study of dose inaccuracies in commercial dose algorithmsMedical Physics, 2013
- Effect of spine hardware on small spinal stereotactic radiosurgery dosimetryPhysics in Medicine & Biology, 2013
- Changes realized from extended bit‐depth and metal artifact reduction in CTMedical Physics, 2013
- Surgical margins in head and neck cancer: A contemporary reviewHead & Neck, 2012
- Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurancePhysics in Medicine & Biology, 2007
- Prototype of a Silicon Nitride Ceramic‐Based Miniplate Osteofixation System for the MidfaceOtolaryngology–Head and Neck Surgery, 2006
- Artifacts in CT: Recognition and AvoidanceRadioGraphics, 2004