The Use of Masseter Electromyography with Electroneurography in the Evaluation of Facial Paralysis
Authors:
- 1 September 1990
- journal article
- research article
- Published by Wiley in Otolaryngology–Head and Neck Surgery
- Vol. 103 (3) , 391-395
Abstract
In the evaluation of facial paralysis, electroneurography provides a quantitative assessment of the degree of degeneration. Near the endpoint of degeneration, the facial musculature compound action potential may be contaminated by masseter artifact resulting from supramaximal stimulation. We propose the use of both masseter electromyography and electroneurography in the evaluation of advanced degeneration to better define the current level for supramaximal stimulation, to eliminate masseter artifact, and to more accurately assess the degree of facial nerve degeneration.Keywords
This publication has 5 references indexed in Scilit:
- Facial Electroneurography: Standardization of Electrode PositionJAMA Otolaryngology–Head & Neck Surgery, 1988
- Facial Electroneurography: Clinical and Experimental InvestigationsOtolaryngology–Head and Neck Surgery, 1985
- Clinical electroneurography: Statistical analysis of controlled measurements in twenty‐two normal subjectsThe Laryngoscope, 1981
- Electroneurography in the Prognostication of Bell's PalsyActa Oto-Laryngologica, 1981
- The Acute Facial Palsies1977