Unsuitability and high perforation rate of latex-free gloves in arthroplasty: a cause for concern
Authors:
- 7 July 2010
- journal article
- research article
- Published by Springer Nature in Archives of orthopaedic and trauma surgery
- Vol. 131 (4) , 455-458
Abstract
Introduction Any increase in perforations in surgical gloves is a concern. We conducted a prospective study to find a clinically acceptable brand of latex-free gloves suitable for orthopaedic surgery. Method Five different brands of gloves approved for use were selected. Two of the latex-free gloves were rejected as their clinical usability in arthroplasty was poor. The Cardinal Esteem latex-free glove has a smooth surface and was slippery, especially in total knee arthroplasty. The Biogel Skin Sense under glove was found to be unacceptably thick, affecting both surgical dexterity and tactility. Results The three remaining gloves were Biogel latex (control), Synthesis Polyco (latex-free) and Cardinal Esteem SMT (latex-free). Gloves were then collected for testing from 241 arthroplasties. Gloves used for primary knee/hip replacements were tested using a validated water distension test. Latex-free gloves were found to have higher perforation rates compared to latex. Conclusion Our study demonstrates that currently available latex-free gloves have inferior clinical performance and are not recommended for use in arthroplasty surgery.Keywords
This publication has 12 references indexed in Scilit:
- Latex-Free GlovesThe Journal of Arthroplasty, 2010
- Glove perforation and contamination in primary total hip arthroplastyThe Journal of Bone and Joint Surgery. British volume, 2005
- Health Care Workers: Risk Factors for Nonlatex and Latex Gloves During SurgeryAIHA Journal, 2003
- Randomized trial of blunt-tipped versus cutting needles to reduce glove puncture during mass closure of the abdomenBritish Journal of Surgery, 1996
- An Epidemic of Occupational Allergy to Latex Involving Health Care WorkersJournal of Occupational and Environmental Medicine, 1995
- Prevalence of Latex Sensitization among Hospital Physicians Occupationally Exposed to Latex GlovesAnesthesiology, 1992
- Glove perforation in hand surgeryThe Journal of Hand Surgery, 1991
- Glove punctures in an orthopaedic trauma unitInjury, 1990
- HTLV-III infection among health care workers. Association with needle-stick injuriesJAMA, 1985
- The Risk of Hepatitis B Infection Among Health Professionals in the United States: A ReviewThe American Journal of the Medical Sciences, 1984