Hydroxyapatite (HA) coating appears to be of benefit for implant durability of tibial components in primary total knee arthroplasty
Authors:
- Jeffrey D VoigtJVJeffrey D Voigt
- Michael MosierMMMichael Mosier
Open Access
- 10 June 2011
- journal article
- review article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 82 (4) , 448-459
Abstract
It is unclear whether there is a clinical benefit to adding hydroxyapatite (HA) coatings to total knee implants, especially with the tibial component, where failure of the implant more often occurs. A systematic review of the literature was undertaken to identify all prospective randomized trials for determining whether the overall clinical results (as a function of durability, function, and adverse events) favored HA-coated tibial components. A comprehensive literature search was performed for the years 1990 to September 16, 2010. We restricted our search to randomized controlled trials involving participants receiving either an HA-coated tibia or other forms of tibial fixation. The primary outcome measures evaluated were durability, function, and acute adverse events. Data from 926 evaluable primary total knee implants in 14 studies were analyzed. Using an RSA definition for durability, HA-coated tibial components (porous or press-fit) without screw fixation were less likely to be unstable at 2 years than porous and cemented metal-backed tibial components (RR = 0.58, 95% CI: 0.34-0.98; p = 0.04, I(2) = 39%, M-H random effects model). There was no significant difference in durability, as measured from revision and evaluated at 2 and 8-10 years, between groups. Also, functional status using different validated measures showed no significant difference at 2 and 5 years, no matter what measure was used. Lastly, there was no significant difference in adverse events. Limitations included small numbers of evaluable patients (≤ 50) in 7 of the 14 trials identified, and a lack of "hard" evidence of durability with need for replacement (i.e. frank failure, pain, or loss of functionality). In patients > 65 years of age, an HA-coated tibial implant may provide better durability than other forms of tibial fixation. Larger trials should be undertaken comparing the long-term durability, function, and adverse events of HA-coated implants with those of other porous-coated tibial implants in younger, more active OA patients.Keywords
This publication has 33 references indexed in Scilit:
- Small study effects in meta-analyses of osteoarthritis trials: meta-epidemiological studyBMJ, 2010
- Effect of bioactive coating of the tibial component on migration pattern in uncemented total knee arthroplasty: a randomized RSA study of 14 knees presented according to new RSA-guidelinesJournal of Orthopaedics and Traumatology, 2008
- A randomised RSA study of Peri-Apatite™ HA coating of a total knee prosthesisThe Knee, 2008
- Cemented versus hydroxyapatite fixation of the femoral component of the Freeman-Samuelson total knee replacementThe Journal of Bone and Joint Surgery. British volume, 2007
- Five- to 12-year follow-up of a hydroxyapatite-coated, cementless total knee replacement in young, active patientsThe Journal of Bone and Joint Surgery. British volume, 2006
- Uncemented HA-coated Implant is the Optimum Fixation for TKA in the Young PatientClinical Orthopaedics and Related Research, 2006
- Periapatite May Not Improve Micromotion of Knee Prostheses in Rheumatoid ArthritisClinical Orthopaedics and Related Research, 2006
- Effect of Hydroxyapatite-Coated Tibial Components on Changes in Bone Mineral Density of the Proximal Tibia After Uncemented Total Knee ArthroplastyThe Journal of Arthroplasty, 2005
- Sealing effect of hydroxyapatite coating on peri-implant migration of particlesThe Journal of Bone and Joint Surgery. British volume, 2001
- Sealing effect of hydroxyapatite coating: A 12-month study in caninesActa Orthopaedica, 2000