The properties of a new commercial thermoluminescent dosimeter, CaSO4:Dy without a LiF binder (TLD‐900) was studied for low‐exposure measurements in diagnostic radiology. The former TLD‐900 had a LiF binder (herein referred to as “TLD‐900/LiF”). The principle features of this dosimeter are its high sensitivity to low‐energy radiation and its relatively low fading which permits measurements down to less than 2.6×10−8 C kg (0.1 mR) with an accuracy better than 20%. The characteristics, annealing procedures, light sensitivity, energy response, reproducibility, and fading, of TLD‐900 are discussed and compared with TLD‐900/LiF. When the precautions presented in this paper are used, the dosimeters can be used for the measurement of x‐ray exposures.