Specific absorption rate benefits of including measured electric field interactions in parallel excitation pulse design
Open Access
- 29 August 2011
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 67 (1) , 164-174
Abstract
Specific absorption rate management and excitation fidelity are key aspects of radiofrequency pulse design for parallel transmission at ultra–high magnetic field strength. The design of radiofrequency pulses for multiple channels is often based on the solution of regularized least‐squares optimization problems for which a regularization term is typically selected to control the integrated or peak pulse waveform amplitude. Unlike single‐channel transmission, the specific absorption rate of parallel transmission is significantly influenced by interferences between the electric fields associated with the individual transmission elements, which a conventional regularization term does not take into account. This work explores the effects upon specific absorption rate of incorporating experimentally measurable electric field interactions into parallel transmission pulse design. Results of numerical simulations and phantom experiments show that the global specific absorption rate during parallel transmission decreases when electric field interactions are incorporated into pulse design optimization. The results also show that knowledge of electric field interactions enables robust prediction of the net power delivered to the sample or subject by parallel radiofrequency pulses before they are played out on a scanner. Magn Reson Med, 2011.Keywords
Funding Information
- NIH (R01-EB011551, R01-EB000447)
This publication has 23 references indexed in Scilit:
- Electrodynamic constraints on homogeneity and radiofrequency power deposition in multiple coil excitationsMagnetic Resonance in Medicine, 2009
- Reducing SAR in parallel excitation using variable-density spirals: a simulation-based studyMagnetic Resonance Imaging, 2008
- Spatial domain method for the design of RF pulses in multicoil parallel excitationMagnetic Resonance in Medicine, 2006
- Fast‐kzthree‐dimensional tailored radiofrequency pulse for reducedB1inhomogeneityMagnetic Resonance in Medicine, 2006
- Basic considerations on the impact of the coil array on the performance of Transmit SENSEMagnetic Resonance Materials in Physics, Biology and Medicine, 2005
- Parallel excitation with an array of transmit coilsMagnetic Resonance in Medicine, 2004
- Transmit SENSEMagnetic Resonance in Medicine, 2002
- Two‐dimensional spatially‐selective RF excitation pulses in echo‐planar imagingMagnetic Resonance in Medicine, 2002
- Three-dimensional tailored RF pulses for the reduction of susceptibility artifacts inT*2-weighted functional MRIMagnetic Resonance in Medicine, 2000
- A k-space analysis of small-tip-angle excitationJournal of Magnetic Resonance (1969), 1989