Repeatability, reproducibility and intersession variability of the Allergan Humphrey ultrasonic biometer
- 1 June 1992
- journal article
- Published by Wiley in Acta Ophthalmologica
- Vol. 70 (3) , 327-334
Abstract
We evaluated repeatability and reproducibility of the Allergan Humphrey model 820 ultrasonic biometer. Sixty eyes (30 subjects) free from ophthalmological abnormality were examined on two separate occasions by two experimenters, one experienced and one inexperienced. The 95% confidence limits for repeatability lie between +0.12/‐0.13 mm and reproducibility between +0.15/‐0.14 mm for AC depth, between +0.11/‐0.13 mm and +0.12/‐0.16 mm for lens thickness, between +0.17/‐0.19 mm and +0.19/‐0.18 mm for right axial length and between +0.26/‐0.40 mm and +0.22/‐0.25 mm for left axial length. The 40% gain used to measure left axial lengths produced poor retinal echospikes, which may account for reduced accuracy compared to the right eyes using 60% gain. Ten subjects (20 eyes) returned for 3 further visits and the variability over 5 sessions was examined using a two‐factor ANOVA with repeated measures on one factor. There were no significant experimenter or session effects and no interaction effect. This biometer gives clinically repeatable and reproducible results and is convenient to use.Keywords
This publication has 24 references indexed in Scilit:
- Effects of different ocular fixation conditions on A-scan ultrasound biometry measurementsOphthalmic and Physiological Optics, 1991
- Calculated versus A-scan result for axial length using different types of ultrasound probe tipEye, 1990
- Immersion versus contact technique in the measurement of axial length by ultrasoundActa Ophthalmologica, 1989
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- ANALYSIS OF ECHOES FROM THE REAR EYE WALL WITH THE AID OF EXPERIMENTAL RESEARCHActa Ophthalmologica, 1964
- POWER OF THE CORRECTING LENS OF THE APHAKIC EYE AS CALCULATED FROM THE KERATOMETRIC MEASUREMENT OF THE CORNEAL RADIUS AND THE ULTRASONICALLY‐MEASURED AXIAL LENGTH OF THE EYEActa Ophthalmologica, 1963
- MEASUREMENT OF INTRAOCULAR DISTANCES BY ULTRASOUND AND COMPARISON BETWEEN OPTICAL AND ULTRASONIC DETERMINATIONS OF THE DEPTH OF THE ANTERIOR CHAMBERActa Ophthalmologica, 1963
- DETERMINATION OF THE VELOCITY OF ULTRASOUND IN THE HUMAN LENS AND VITREOUSActa Ophthalmologica, 1962
- DETERMINATION OF THE VELOCITY OF ULTRASOUND IN OCULAR TISSUES AT DIFFERENT TEMPERATURESActa Ophthalmologica, 1961
- Measurement of the Diameters of the Living Eye by Means of X-RaysNature, 1945