Comparison of the fatigue characteristics of centrifuged and uncentrifuged simplex P bone cement
Authors:
- 1 January 1987
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 5 (3) , 366-371
Abstract
Fatigue test specimens of Simplex P bone cement (Howmedica Inc., Rutherford, NJ) prepared according to the manufacturer's instructions and specimens of Simplex P prepared by centrifuging the cement immediately after mixing were subjected to fully reversed tension‐compression fatigue tests at initial strain levels ranging from 0.01 to 0.001. S:N curves for both cement preparations were generated and compared on the basis of regression analysis, Weibull analysis, and Student's t tests. Centrifuged Simplex P was able to withstand significantly more fatigue cycles than uncentrifuged Simplex P at all strain levels tested. Importantly, at the more physiologic strain levels of 2,000 and 1,000 microstrain, the centrifuged cement demonstrated superior fatigue life. Specifically, at the 1,000‐microstrain level eight of the 11 uncentrifuged specimens fractured before undergoing 10 million cycles. Of those that failed, the average number of cycles to failure was 1.8 million with a range of 560,000 to 4 million cycles. In contrast, all 11 centrifuged specimens tested at 1,000 microstrain remained intact at 10 million cycles.Keywords
This publication has 13 references indexed in Scilit:
- Centrifugation as a method of improving tensile and fatigue properties of acrylic bone cement.Journal of Bone and Joint Surgery, 1984
- Tensile Fatigue Failure of Acrylic Bone CementJournal of Biomechanical Engineering, 1983
- A ten-year follow-up of one hundred consecutive Müller curved-stem total hip-replacement arthroplasties.Journal of Bone and Joint Surgery, 1982
- Ten-year follow-up study of total hip replacement.Journal of Bone and Joint Surgery, 1982
- Strain‐controlled fatigue of acrylic bone cementJournal of Biomedical Materials Research, 1982
- An analysis of femoral component stem design in total hip arthroplasty.Journal of Bone and Joint Surgery, 1980
- Fracture characteristics of acrylic bone cements. II. FatigueJournal of Biomedical Materials Research, 1977
- Mechanical Properties of Carbon Fiber-Reinforced Polymethylmethacrylate for Surgical Implant Applications1977
- Carbon fiber‐reinforced bone cement in orthopedic surgeryJournal of Biomedical Materials Research, 1976
- A Statistical Distribution Function of Wide ApplicabilityJournal of Applied Mechanics, 1951