Abstract
A universal experimental challenge in studying radiation effects on cryogenic devices is precisely and accurately characterizing the position-dependent device response near the energy detection threshold. We have developed a compact cryogenic optical beam steering system that can be used to generate O (μs) pulses of small numbers of photons over the energy range of 1.3–3.4 eV at room temperature, and deliver those photons via fiber optic to any specified location on the surface of a detector operating at cryogenic temperatures. This new system will allow for robust calibration of any photon-sensitive detector, including superconducting devices. The system can be used efficiently to explore the physics of target materials, quantify the position sensitivity of different sensor designs, measure phonon transport, and study the effects of quasiparticle poisoning on detector operation. We describe the design of this pulsed calibration method and present first results obtained with a second-generation system operated at room temperature and sub-Kelvin temperatures.
Funding Information
  • Kavli Institute for Particle Astrophysics and Cosmology
  • Army Research Office (W911NF-22-1-0257)
  • Geoff and Josie Fox Fellowship
  • Lee and Seymour Graff Endowed Chair Fund
  • Office of Science
  • Quantum Science Center, Q-NEXT

This publication has 15 references indexed in Scilit: