Exposure levels of radiofrequency magnetic fields and static magnetic fields in 1.5 and 3.0 T MRI units
Authors:
Open Access
- 19 January 2021
- journal article
- research article
- Published by Springer Nature in SN Applied Sciences
- Vol. 3 (2) , 1-13
Abstract
Magnetic resonance imaging (MRI) staff is exposed to a complex mixture of electromagnetic fields from MRI units. Exposure to these fields results in the development of transient exposure-related symptoms. This study aimed to investigate the exposure levels of radiofrequency (RF) magnetic fields and static magnetic fields (SMFs) from 1.5 and 3.0 T MRI scanners in two public hospitals in the Mangaung Metropolitan region, South Africa. The exposure levels of SMFs and RF magnetic fields were measured using the THM1176 3-Axis hall magnetometer and TM-196 3 Axis RF field strength meter, respectively. Measurements were collected at a distance of 1 m (m) and 2 m from the gantry for SMFs when the brain, cervical spine and extremities were scanned. Measurements for RF magnetic fields were collected at a distance of 1 m with an average scan duration of six minutes. Friedman’s test was used to compared exposure mean values from two 1.5 T scanners, and Wilcoxon test with Bonferroni adjustment was used to identify where the difference between exist. The Shapiro–Wilk test was also used to test for normality between exposure levels in 1.5 and 3.0 T scanners. The measured peak values for SMFs from the 3.0 T scanner at hospital A were 1300 milliTesla (mT) and 726 mT from 1.5 T scanner in hospital B. The difference in terms of SMFs exposure levels was observed between two 1.5 T scanners at a distance of 2 m. The difference between 1.5 T scanners at 1 m was also observed during repeated measurements when brain, cervical spine and extremities scans were performed. Scanners’ configurations, magnet type, clinical setting and location were identified as factors that could influence different propagation of SMFs between scanners of the same nominal B0. The RF pulse design, sequence setting flip-angle and scans performed influenced the measured RF magnetic fields. Three scanners were complaint with occupational exposure guidelines stipulated by the ICNIRP; however, peak levels that exist at 1 m could be managed through adoption of occupational health and safety programs.Keywords
This publication has 36 references indexed in Scilit:
- Inventory of MRI applications and workers exposed to MRI-related electromagnetic fields in the NetherlandsEuropean Journal of Radiology, 2013
- Cognition and Sensation in Very High Static Magnetic Fields: A Randomized Case-Crossover Study with Different Field StrengthsRadiology, 2013
- Exposure classification of MRI workers in epidemiological studiesBioelectromagnetics, 2012
- Occupational exposure in MRIBritish Journal of Radiology, 2012
- Health complaints among nurses working near MRI scanners—A descriptive pilot studyEuropean Journal of Radiology, 2011
- MR safety: simultaneous B 0, dΦ/dt, and dB/dt measurements on MR-workers up to 7 TMagnetic Resonance Materials in Physics, Biology and Medicine, 2011
- GUIDELINES ON LIMITS OF EXPOSURE TO STATIC MAGNETIC FIELDSHealth Physics, 2009
- Occupational risk from static magnetic fields of MRI scannersEnvironmentalist, 2007
- EU Directive 2004/40: field measurements of a 1.5 T clinical MR scannerBritish Journal of Radiology, 2007
- Occupational Exposure Assessment of the Static Magnetic Flux Density Generated by Nuclear Magnetic Resonance Spectroscopy for Biochemical PurposesPIERS Online, 2007