The Effect of Dehydration on Non-Invasive Hemoglobin Values [Masimo®] in Adult Males—An Exploratory Cross-Sectional Study
Open Access
- 7 May 2026
- journal article
- research article
- Published by MDPI AG in Emergency Care and Medicine
- Vol. 3 (2) , 18
Abstract
Background: Non-invasive, real-time hemoglobin monitoring (SpHb) may reduce blood-draw costs and accelerate clinical decision-making. This study evaluated agreement between SpHb (Masimo®) and laboratory hemoglobin (LabHb) in patients with and without dehydration. Methods/Approach: This single-center exploratory cross-sectional study included male veterans. SpHb and LabHb were measured simultaneously. Demographic data, fasting status, and laboratory markers of dehydration were extracted from the electronic health record. Descriptive statistics, correlation analyses, and Bland–Altman analyses were performed. Results: Fifty-three male veterans were included. LabHb ranged from 9.20–17.00 g·dL−1 and SpHb ranged from 9.90–15.90 g·dL−1. Patients with dehydration had higher LabHb (M = 14.54, SD = 1.60) than those without dehydration (M = 13.01, SD = 1.68; p < 0.001). SpHb and LabHb were strongly correlated in non-dehydrated patients (r = 0.82, p < 0.001) but not in dehydrated patients (r = 0.23, p = 0.275). Among non-dehydrated patients, limits of agreement were −1.67 to 2.18 g·dL−1 (~69% within threshold). In dehydrated patients, limits widened to −1.97 to 5.27 g·dL−1 (~25% within threshold), with 70.83% overestimating LabHb (>1 g·dL−1). Conclusion: This exploratory study found a strong correlation and acceptable agreement between SpHb and LabHb in non-dehydrated patients, with substantially reduced agreement in dehydrated patients. Only approximately 25% of measurements in dehydrated patients met predefined agreement limits, indicating clinically meaningful variability. These findings suggest that hydration status may significantly affect SpHb performance. While SpHb may be useful in appropriately selected, hydrated populations, caution is warranted in dehydrated states. Larger, adequately powered studies are needed to further evaluate the impact of hydration and define the optimal patient populations and clinical applications.Keywords
This publication has 20 references indexed in Scilit:
- Influence of optical clearing agents on the scattering properties of human nail bed and blood microrheological properties: In vivo and in vitro studyJournal of Biophotonics, 2024
- Accuracy of non-invasive hemoglobin level measurement in the emergency departmentThe American Journal of Emergency Medicine, 2022
- Evaluation of the usefulness of non-invasive serum haemoglobin measurement in a perioperative setting in a prospective observational studyScientific Reports, 2022
- The Comparison of Hemoglobin Values Measured by Blood and Continuous Non-Invasive Monitoring (SpHb) in Newborn InfantsJournal of Tropical Pediatrics, 2020
- Continuous hemoglobin and plethysmography variability index monitoring can modify blood transfusion practice and is associated with lower mortalityJournal of Clinical Monitoring and Computing, 2019
- Accuracy and trending of non-invasive hemoglobin measurement during different volume and perfusion statusesJournal of Clinical Monitoring and Computing, 2018
- Prevalence and factors predictive of full stomach in elective and emergency surgical patients: a prospective cohort studyBritish Journal of Anaesthesia, 2017
- Evaluation of Noninvasive Hemoglobin Monitoring in Surgical Critical Care Patients*Critical Care Medicine, 2016
- The measurement of dyshemoglobins and total hemoglobin by pulse oximetryCurrent Opinion in Anaesthesiology, 2008
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986