Safety and efficacy of a fully closed-loop control ventilation (IntelliVent-ASV®) in sedated ICU patients with acute respiratory failure: a prospective randomized crossover study
- 30 March 2012
- journal article
- research article
- Published by Springer Nature in Intensive Care Medicine
- Vol. 38 (5) , 781-787
Abstract
IntelliVent-ASV® is a development of adaptive support ventilation (ASV) that automatically adjusts ventilation and oxygenation parameters. This study assessed the safety and efficacy of IntelliVent-ASV® in sedated intensive care unit (ICU) patients with acute respiratory failure. This prospective randomized crossover comparative study was conducted in a 12-bed ICU in a general hospital. Two periods of 2 h of ventilation in randomly applied ASV or IntelliVent-ASV® were compared in 50 sedated, passively ventilated patients. Tidal volume (V T), respiratory rate (RR), inspiratory pressure (P INSP), SpO2 and ETCO2 were continuously monitored and recorded breath by breath. Mean values over the 2-h period were calculated. Respiratory mechanics, plateau pressure (P PLAT) and blood gas exchanges were measured at the end of each period. There was no safety issue requiring premature interruption of IntelliVent-ASV®. Minute ventilation (MV) and V T decreased from 7.6 (6.5–9.5) to 6.8 (6.0–8.0) L/min (p < 0.001) and from 8.3 (7.8–9.0) to 8.1 (7.7–8.6) mL/kg PBW (p = 0.003) during IntelliVent-ASV® as compared to ASV. P PLAT and FiO2 decreased from 24 (20–29) to 20 (19–25) cmH2O (p = 0.005) and from 40 (30–50) to 30 (30–39) % (p < 0.001) during IntelliVent-ASV® as compared to ASV. RR, P INSP, and PEEP decreased as well during IntelliVent-ASV® as compared to ASV. Respiratory mechanics, pH, PaO2 and PaO2/FiO2 ratio were not different but PaCO2 was slightly higher during IntelliVent-ASV® as compared to ASV. In passive patients with acute respiratory failure, IntelliVent-ASV® was safe and able to ventilate patients with less pressure, volume and FiO2 while producing the same results in terms of oxygenation.Keywords
This publication has 23 references indexed in Scilit:
- Effects of Implementing Adaptive Support Ventilation in a Medical Intensive Care UnitRespiratory Care, 2011
- Adaptive support ventilation for faster weaning in COPD: a randomised controlled trialEuropean Respiratory Journal, 2011
- Randomized Controlled Trial Comparing Adaptive-support Ventilation with Pressure-regulated Volume-controlled Ventilation with Automode in Weaning Patients after Cardiac SurgeryAnesthesiology, 2008
- Burnout syndrome among critical care healthcare workersCurrent Opinion in Critical Care, 2007
- Automatic selection of breathing pattern using adaptive support ventilationIntensive Care Medicine, 2007
- Closed-Loop Ventilation: An Emerging Standard of Care?Critical Care Clinics, 2007
- Higher versus Lower Positive End-Expiratory Pressures in Patients with the Acute Respiratory Distress SyndromeNew England Journal of Medicine, 2004
- Hospital mortality in relation to staff workload: a 4-year study in an adult intensive-care unitThe Lancet, 2000
- Respiratory mechanics by least squares fitting in mechanically ventilated patients: Applications during paralysis and during pressure support ventilationIntensive Care Medicine, 1995
- The American-European Consensus Conference On Ards. Definitions, Mechanisms, Relevant Outcomes, And Clinical Trial CoordinationAmerican Journal of Respiratory and Critical Care Medicine, 1994