Abstract
Background and Aims: A preliminary study compared the use of continuous glucose monitoring (CGM) with the use of self-monitored blood glucose (SMBG) by aircraft pilots with insulin-treated diabetes in the United Kingdom, Ireland, and Austria, certified to fly commercial aircraft within the European Aviation Safety Agency ARA.MED.330 protocol. Methods: SMBG and simultaneous interstitial glucose measurements using CGM (Dexcom G6 ® ) were recorded during pre- and in-flight periods. Results: Eight male pilots (seven with type 1 diabetes and one with type 3c diabetes), median age of 48.5 years and median diabetes duration of 11.5 years, participated. The correlation coefficient ( R ) between 874 contemporaneously recorded SMBG and CGM values was 0.843, P < 0.001. The mean glucose concentration was 8.78 mmol/L (standard deviation [SD] 0.67) using SMBG compared with 8.71 mmol/L (SD 0.85) recorded using CGM. The mean absolute relative difference was 9.39% (SD 3.12). Conclusions: CGM using Dexcom G6 systems is a credible alternative to SMBG for monitoring glucose levels when insulin-treated pilots fly commercial aircraft. The study was registered with Clinical Trials.gov NCT04395378.

This publication has 32 references indexed in Scilit: