Microdosimetric measurements of a monoenergetic and modulated Bragg Peaks of 62 MeV therapeutic proton beam with a synthetic single crystal diamond microdosimeter
Authors:
- 24 September 2020
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 47 (11) , 5791-5801
Abstract
Purpose: The purpose of this study was to investigate for the first time the performance of a synthetic single crystal diamond detector for the microdosimetric characterization of clinical 62 MeV ocular therapy proton beams.Methods: A novel diamond microdosimeter with a well‐defined sensitive volume was fabricated and tested with a monoenergetic and spread‐out Bragg peak (SOBP) of the CATANA therapeutic proton beam in Catania, Italy. The whole sensitive volume of the detector has an active planar‐sectional area of 100 µm × 100 µm and a thickness of approximately 6.3 um. Microdosimetric measurements were performed at several water equivalent depths, corresponding to positions of clinical relevance. From the measured spectra, microdosimetric quantities such as the frequency mean lineal energy (), dose mean lineal energy () as well as microdosimetric relative biological effectiveness (RBEµ) values were derived for each depth along both a pristine Bragg curve and SOBP. Finally, Geant4 Monte Carlo simulations were performed modeling the detector geometry and CATANA beamline in order to calculate the average linear energy transfer (LET) values in the diamond active layer and water.Results: The microdosimetric spectra acquired by the diamond microdosimeter show different shapes as a function of the water equivalent depths. No spectral distortion, due to pile‐up events and polarization effects, was observed. The experimental spectra have a very low detection threshold due to the electronic noise during the irradiation of about 1 keV/μm. The and values were in agreement with expected trends, showing a sharp increase in mean lineal energy at the distal edge of the Bragg peak. In addition, a good agreement between the mean lineal energy values and the calculated average LET ones was also observed. Finally, the RBE values evaluated with the diamond microdosimeter were in excellent agreement with those obtained with a mini tissue equivalent proportional counter as well as with radiobiological measurements in the same proton beam field.Conclusions: The microdosimetric performance of the tested synthetic single crystal diamond microdosimeter clearly indicates its suitability for quality assurance in clinical proton therapy beam.Keywords
This publication has 48 references indexed in Scilit:
- Microdosimetric calculation of relative biological effectiveness for design of therapeutic proton beamsJournal of Radiation Research, 2012
- In vitro RBE-LET dependence for multiple particle typesActa Oncologica, 2011
- Heavy-ion tumor therapy: Physical and radiobiological benefitsReviews of Modern Physics, 2010
- Charged particles in radiation oncologyNature Reviews Clinical Oncology, 2009
- RBE estimation of proton radiation fields using a telescopeMedical Physics, 2009
- Current Status and New Developments in Ion TherapyStrahlentherapie und Onkologie, 2007
- Microdosimetric investigation at the therapeutic proton beam facility of CATANARadiation Protection Dosimetry, 2004
- Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water; Technical Reports Series No. 398,Health Physics, 2001
- Microdosimetry spectra of the Loma Linda proton beam and relative biological effectiveness comparisonsMedical Physics, 1997
- Calculation of relative biological effectiveness for proton beams using biological weighting functionsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1997