Phase Correction of Fourier Transform Ion Cyclotron Resonance Mass Spectra Using MatLab
- 1 January 2011
- journal article
- research article
- Published by American Chemical Society (ACS) in Journal of the American Society for Mass Spectrometry
- Vol. 22 (1) , 138-147
Abstract
FT-ICR mass spectrometry has been limited to magnitude mode for almost 40 years due to the data processing methods used. However, it is well known that phase correction of the data can theoretically produce an absorption-mode spectrum with a mass-resolving power that is as much as twice as high as conventional magnitude mode, and that it also improves the quality of the peak shape. Temporally dispersed frequency sweep excitation followed by a time delay before detection results in a steep quadratic variation in the signal phase with frequency. Viewing this, it is possible to find the correct phase function by performing a quadratic least squares fit, modified by iterating through phase cycles until the correct quadratic function is found. Here, we present a robust manual method to rotate these signals mathematically and generate a “phased” absorption-mode spectrum. The method can, in principle, be automated. Baseline correction is also included to eliminate the accompanying baseline drift. The resulting experimental FT-ICR absorption-mode spectra exhibit a resolving power that is at least 50% higher than that of the magnitude mode.Keywords
This publication has 39 references indexed in Scilit:
- Fourier transform ion cyclotron resonance mass spectrometry: A primerMass Spectrometry Reviews, 1998
- Stored waveform inverse Fourier transform (SWIFT) ion excitation in trapped-ion mass spectometry: Theory and applicationsInternational Journal of Mass Spectrometry and Ion Processes, 1996
- Stored waveform inverse Fourier transform axial excitation/ejection for quadrupole ion trap mass spectrometryAnalytical Chemistry, 1993
- Laboratory-frame and rotating-frame ion trajectories in ion cyclotron resonance mass spectrometryInternational Journal of Mass Spectrometry and Ion Processes, 1990
- Frequency (mass) errors and phase dependence in magnitude-mode apodized Fourier transform—ion cyclotron resonance spectraInternational Journal of Mass Spectrometry and Ion Processes, 1989
- Experimental evaluation of apodization functions for quantitative fourier transform mass spectrometryInternational Journal of Mass Spectrometry and Ion Processes, 1989
- Apodization of FT-ICR spectraInternational Journal of Mass Spectrometry and Ion Processes, 1987
- Dispersion versus absorption (DISPA) plots as an index of static magnetic field inhomogeneity. Use for adjustment of spinning shims in a superconducting magnetJournal of Magnetic Resonance (1969), 1986
- Developments in analytical fourier-transform mass spectrometryAnalytica Chimica Acta, 1985
- Experimental determination of the effects of space charge on ion cyclotron resonance frequenciesInternational Journal of Mass Spectrometry and Ion Processes, 1983