Time‐interleaved acquisition of modes: An analysis of SAR and image contrast implications
Open Access
- 19 August 2011
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 67 (4) , 1033-1041
Abstract
As the magnetic field strength and therefore the operational frequency in MRI are increased, the radiofrequency wavelength approaches the size of the human head/body, resulting in wave effects which cause signal decreases and dropouts. Especially, whole‐body imaging at 7 T and higher is therefore challenging. Recently, an acquisition scheme called time‐interleaved acquisition of modes has been proposed to tackle the inhomogeneity problems in high‐field MRI. The basic premise is to excite two (or more) different B modes using static radiofrequency shimming in an interleaved acquisition, where the complementary radiofrequency patterns of the two modes can be exploited to improve overall signal homogeneity. In this work, the impact of time‐interleaved acquisition of mode on image contrast as well as on time‐averaged specific absorption rate is addressed in detail. Time‐interleaved acquisition of mode is superior in B homogeneity compared with conventional radiofrequency shimming while being highly specific absorption rate efficient. Time‐interleaved acquisition of modes can enable almost homogeneous high‐field imaging throughout the entire field of view in PD, T2, and T2*‐weighted imaging and, if a specified homogeneity criterion is met, in T1‐weighted imaging as well. Magn Reson Med, 2011.Keywords
This publication has 12 references indexed in Scilit:
- RF excitation using time interleaved acquisition of modes (TIAMO) to address B1 inhomogeneity in high‐field MRIMagnetic Resonance in Medicine, 2010
- The Virtual Family—development of surface-based anatomical models of two adults and two children for dosimetric simulationsPhysics in Medicine & Biology, 2009
- Whole‐body imaging at 7T: Preliminary resultsMagnetic Resonance in Medicine, 2008
- Local B1+ shimming for prostate imaging with transceiver arrays at 7T based on subject‐dependent transmit phase measurementsMagnetic Resonance in Medicine, 2008
- Exploring the limits of RF shimming for high‐field MRI of the human headMagnetic Resonance in Medicine, 2006
- Spatial domain method for the design of RF pulses in multicoil parallel excitationMagnetic Resonance in Medicine, 2006
- Combination of optimized transmit arrays and some receive array reconstruction methods can yield homogeneous images at very high frequenciesMagnetic Resonance in Medicine, 2005
- B1 destructive interferences and spatial phase patterns at 7 T with a head transceiver array coilMagnetic Resonance in Medicine, 2005
- Transmit SENSEMagnetic Resonance in Medicine, 2002
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002