Influence of three mechanical bileaflet prosthetic valve designs on the three-dimensional flow field inside a simulated aorta
- 7 October 2010
- journal article
- research article
- Published by Springer Nature in Journal of Artificial Organs
- Vol. 13 (4) , 207-217
Abstract
The current design of the bileaflet valve, the leaflets of which open outside first, differs significantly from the natural valve whose leaflets open center first. This difference generates a completely different flow field in the bileaflet valve compared to that in the natural heart valve. In a previous study, it was demonstrated that the valve design greatly affects the aortic flow field as well as the circulatory flow inside sinuses of Valsalva, using saline solution as a working fluid. A limited discussion on the turbulence flow field that could be generated by the valve was provided. In this continuation of that study, therefore, a dynamic PIV study was conducted to analyze the influence of the heart valve design on the aortic flow field, and particularly on the turbulent profile. This study also aimed to determine the influence of the viscosity of the testing fluid. Three bileaflet prostheses—the St. Jude Medical (SJM), the On-X, and the MIRA valves—were tested under pulsatile flow conditions. Flow through the central orifice of the SJM valve was slower than that through the newer designs. The newer designs tend to show strong flow through all orifices. The On-X valve generates simple jet-type flow while the MIRA valve with circumferentially curved leaflets generates a strong but three-dimensionally diffuse flow, resulting in a more complex flow field downstream of the aortic valve with higher turbulence. A 180° orientation that is more popular clinically seems to provide a less diffuse flow than a 90° orientation. The effect of increasing the viscosity was found to be an increase in the flow velocity through the central orifice and a more organized flow field for all of the valves tested.Keywords
This publication has 10 references indexed in Scilit:
- Dynamic particle image velocimetry study of the aortic flow field of contemporary mechanical bileaflet prosthesesJournal of Artificial Organs, 2008
- 129 Dynamic PIV flow field analysis of the mechanical bi-leaflet valves : Influence of working fluid viscosityThe Proceedings of the Bioengineering Conference Annual Meeting of BED/JSME, 2008
- Two-Dimensional Dynamic Simulation of Platelet Activation During Mechanical Heart Valve ClosureAnnals of Biomedical Engineering, 2006
- Dynamic particle image velocimetry flow analysis of the flow field immediately downstream of bileaflet mechanical mitral prosthesesJournal of Artificial Organs, 2006
- Time-resolved particle image velocimetry and laser doppler anemometry study of the turbulent flow field of bileaflet mechanical mitral prosthesesJournal of Artificial Organs, 2005
- Unsteady Fluid Dynamics of Several Mechanical Prosthetic Heart Valves Using a Two Component Laser Doppler Anemometer SystemArtificial Organs, 1997
- Shear-induced activation of plateletsJournal of Biomechanics, 1979
- The Anatomy of the Mitral Valve and its Associated StructuresThorax, 1964
- Aortic ReplacementCirculation, 1963
- Surgical correction of aortic insufficiency.1954