Relationship between Noise, Dose, and Pitch in Cardiac Multi–Detector Row CT

Authors:
  • CMCynthia H. McCollough
    Cynthia H. McCollough
  • MBMichael R. Bruesewitz
    Michael R. Bruesewitz
  • JZJie Zhang
    Jie Zhang
  • JFJoel G. Fletcher
    Joel G. Fletcher

Abstract
In spiral computed tomography (CT), dose is always inversely proportional to pitch. However, the relationship between noise and pitch (and hence noise and dose) depends on the scanner type (single vs multi–detector row) and reconstruction mode (cardiac vs noncardiac). In single detector row spiral CT, noise is independent of pitch. Conversely, in noncardiac multi–detector row CT, noise depends on pitch because the spiral interpolation algorithm makes use of redundant data from different detector rows to decrease noise for pitch values less than 1 (and increase noise for pitch values > 1). However, in cardiac spiral CT, redundant data cannot be used because such data averaging would degrade the temporal resolution. Therefore, the behavior of noise versus pitch returns to the single detector row paradigm, with noise being independent of pitch. Consequently, since faster rotation times require lower pitch values in cardiac multi–detector row CT, dose is increased without a commensurate decrease in noise. Thu...

This publication has 22 references indexed in Scilit: