Physiological effects of human body imaging with 300 mT/m gradients
Open Access
- 21 December 2021
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 87 (5) , 2512-2520
Abstract
Purpose: The use of high‐performance gradient systems (i.e., high gradient strength and/or high slew rate) for human MRI is limited by physiological effects (including the elicitation of magnetophosphenes and peripheral nerve stimulation (PNS)). These effects, in turn, depend on the interaction between time‐varying magnetic fields and the body, and thus on the participant’s position with respect to the scanner’s isocenter. This study investigated the occurrence of magnetophosphenes and PNS when scanning participants on a high‐gradient (300 mT/m) system, for different gradient amplitudes, ramp times, and participant positions.Methods: Using a whole‐body 300 mT/m gradient MRI system, a cohort of participants was scanned with the head, heart, and prostate at magnet isocenter and a train of trapezoidal bipolar gradient pulses, with ramp times from 0.88 to 4.20 ms and gradient amplitudes from 60 to 300 mT/m. Reports of magnetophosphenes and incidental reports of PNS were obtained. A questionnaire was used to record any additional subjective effects.Results: Magnetophosphenes were strongly dependent on participant position in the scanner. 87% of participants reported the effect with the heart at isocenter, 33% with the head at isocenter, and only 7% with the prostate at isocenter. PNS was most widely reported by participants for the vertical gradient axis (67% of participants), and was the dominant physiological effect for ramp times below 2 ms.Conclusion: This study evaluates the probability of eliciting magnetophosphenes during whole‐body imaging using an ultra‐strong gradient MRI system. It provides empirical guidance on the use of high‐performance gradient systems for whole‐body human MRI.Keywords
Funding Information
- Nederlandse Organisatie voor Wetenschappelijk Onderzoek (17331)
- Engineering and Physical Sciences Research Council (EP/M029778/1)
This publication has 43 references indexed in Scilit:
- Retinal origin of phosphenes to transcranial alternating current stimulationClinical Neurophysiology, 2010
- Retina or visual cortex? The site of phosphene induction by transcranial alternating current stimulationFrontiers in Integrative Neuroscience, 2009
- Stimulation threshold comparison of time‐varying magnetic pulses with different waveformsJournal of Magnetic Resonance Imaging, 2008
- Avoiding peripheral nerve stimulation: Gradient waveform criteria for optimum resolution in echo‐planar imagingMagnetic Resonance in Medicine, 1994
- Limits to neural stimulation in echo‐planar imagingMagnetic Resonance in Medicine, 1993
- Health and Physiological Effects of Human Exposure to Whole‐Body Four‐Tesla Magnetic Fields during MRIAnnals of the New York Academy of Sciences, 1992
- Magnetostimulation of Vision: Direct Noninvasive Stimulation of the Retina and the Visual BrainOptometry and Vision Science, 1991
- On the history of deformation phosphenes and the idea of internal light generated in the eye for the purpose of visionDocumenta Ophthalmologica, 1990
- On the history of deformation phosphenes and the idea of internal light generated in the eye for the purpose of vision1990
- Stimulation of cardiac muscle by a time-varying magnetic fieldIEEE Transactions on Magnetics, 1970