Implementation of Size-Dependent Local Diagnostic Reference Levels for CT Angiography

Authors:

Abstract
OBJECTIVE. Diagnostic reference levels (DRLs) are established for standard-sized patients; however, patient dose in CT depends on patient size. The purpose of this study was to introduce a method for setting size-dependent local diagnostic reference levels (LDRLs) and to evaluate these LDRLs in comparison with size-independent LDRLs and with respect to image quality. MATERIALS AND METHODS. One hundred eighty-four aortic CT angiography (CTA) examinations performed on either a second-generation or third-generation dual-source CT scanner were included; we refer to the second-generation dual-source CT scanner as “CT1” and the third-generation dual-source CT scanner as “CT2.” The volume CT dose index (CTDIvol) and patient diameter (i.e., the water-equivalent diameter) were retrieved by dose-monitoring software. Size-dependent DRLs based on a linear regression of the CTDIvol versus patient size were set by scanner type. Size-independent DRLs were set by the 5th and 95th percentiles of the CTDIvol values. Object...

This publication has 13 references indexed in Scilit: