Abstract
The modern cystoscope is the result of the advancement in technology in numerous areas and is an invaluable tool that allows the urologist to perform a number of diagnostic and therapeutic procedures. Although various degrees of endoscope failure have been widely reported, instrument breakage that leads to a foreign body has not. While performing a bladder neck stricture incision for a 72-year-old male patient with a previous radical prostatectomy for prostate cancer and a resulting bladder neck stricture, we documented a major 17-French cystoscope malfunction and a resulting foreign body that was retrieved from the bladder using a 22-French scope and alligator forceps.

This publication has 11 references indexed in Scilit: