Measurement of radiation damping rate constants in nuclear magnetic resonance by inversion recovery and automated compensation of selective pulses
- 15 April 2000
- journal article
- research article
- Published by AIP Publishing in The Journal of Chemical Physics
- Vol. 112 (15) , 6511-6514
Abstract
A method is proposed to measure the radiation damping rate constant based on the analysis of the nonexponential recovery of the magnetization after inversion. It is applicable when the recovery is dominated by radiation damping rather than by relaxation processes. The accurate measurement of the radiation damping rate constant can be used to simplify a recent method for the compensation of radiation damping effects during the application of selective pulses [Chen, Jerschow, and Bodenhausen, Chem. Phys. Lett. 308, 397 (1999)]. This procedure can now be automated, as illustrated by applications to selective inversion and excitation.Keywords
This publication has 7 references indexed in Scilit:
- Compensation of radiation damping during selective pulses in NMR spectroscopyChemical Physics Letters, 1999
- Experimental verification of the relation between the delay time and the tipping angle for a superradiant pulseJournal of Physics D: Applied Physics, 1999
- Line shapes of strongly radiation-damped nuclear magnetic resonance signalsThe Journal of Chemical Physics, 1993
- Dynamics of radiation damping in nuclear magnetic resonanceThe Journal of Chemical Physics, 1989
- Radiation Damping in Nuclear Magnetic ResonancePhysical Review B, 1959
- Effects of Radiation Damping on Spin DynamicsJournal of Applied Physics, 1957
- Radiation Damping in Magnetic Resonance ExperimentsPhysical Review B, 1954