3D-Printed Biomaterial Testing in Response to Cryoablation: Implications for Surgical Ventricular Tachycardia Ablation
Authors:
- Mara CandelariMCMara Candelari
- Ida Anna CappelloICIda Anna Cappello
- Luigi PannoneLPLuigi Pannone
- Cinzia MonacoCMCinzia Monaco
- Edoardo BoriEBEdoardo Bori
- Giacomo TaleviGTGiacomo Talevi
- Robbert RamakRRRobbert Ramak
- Mark La MeirMMMark La Meir
- Ali GharaviriAGAli Gharaviri
- Gian Battista ChierchiaGCGian Battista Chierchia
- Carlo de AsmundisCarlo de Asmundis
- Bernardo InnocentiBernardo Innocenti
Open Access
- 29 January 2023
- journal article
- research article
- Published by MDPI AG in Journal of Clinical Medicine
- Vol. 12 (3) , 1036
Abstract
Background: The lack of thermally and mechanically performant biomaterials represents the major limit for 3D-printed surgical guides, aimed at facilitating complex surgery and ablations. Methods: Cryosurgery is a treatment for cardiac arrhythmias. It consists of obtaining cryolesions, by freezing the target tissue, resulting in selective and irreversible damage. MED625FLX and TPU95A are two biocompatible materials for surgical guides; however, there are no data on their response to cryoenergy delivery. The study purpose is to evaluate the biomaterials’ thermal properties, examining the temperature changes on the porcine muscle samples (PMS) when the biomaterials are in place during the cryoablation. Two biomaterials were selected, MED625FLX and TPU95A, with two thicknesses (1.0 and 2.5 mm). To analyze the biomaterials’ behavior, the PMS temperatures were measured during cryoablation, firstly without biomaterials (control) and after with the biomaterials in place. To verify the biomaterials’ suitability, the temperatures under the biomaterial samples should not exceed a limit of −30.0 °C. Furthermore, the biomaterials’ geometry after cryoablation was evaluated using the grid paper test. Results: TPU95A (1.0 and 2.5 mm) successfully passed all tests, making this material suitable for cryoablation treatment. MED625FLX of 1.0 mm did not retain its shape, losing its function according to the grid paper test. Further, MED625FLX of 2.5 mm is also suitable for use with a cryoenergy source. Conclusions: TPU95A (1.0 and 2.5 mm) and MED625FLX of 2.5 mm could be used in the design of surgical guides for cryoablation treatment, because of their mechanical, geometrical, and thermal properties. The positive results from the thermal tests on these materials and their thickness prompt further clinical investigation.Keywords
This publication has 19 references indexed in Scilit:
- Cryosurgery for Atrial Fibrillation: Physiologic Basis for Creating Optimal CryolesionsThe Annals of Thoracic Surgery, 2021
- Cryoablation: physical and molecular basis with putative immunological consequencesInternational Journal of Hyperthermia, 2019
- Fabrication and Characterization of Electrospun Thermoplastic Polyurethane/Fibroin Small-Diameter Vascular Grafts for Vascular Tissue EngineeringInternational Polymer Processing, 2016
- Evaluation of a novel cryoablation system: in vivo testing in a chronic porcine model.2013
- Catheter CryoablationCirculation: Arrhythmia and Electrophysiology, 2013
- Evaluation of a Novel Cryoablation System: In vivo Testing in a Chronic Porcine ModelInnovations: Technology and Techniques in Cardiothoracic and Vascular Surgery, 2012
- Normal Left Ventricular Myocardial Thickness for Middle-Aged and Older Subjects With Steady-State Free Precession Cardiac Magnetic ResonanceCirculation: Cardiovascular Imaging, 2012
- Experimental cryosurgery investigations in vivoCryobiology, 2009
- Mechanisms of Tissue Injury in CryosurgeryCryobiology, 1998
- Cardiac cryosurgery: effects of myocardial temperature on cryolesion size.1983