Performance of Pyrolyzed Photoresist Carbon Films in a Microchip Capillary Electrophoresis Device with Sinusoidal Voltammetric Detection
- 11 July 2003
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 75 (16) , 4265-4271
Abstract
Pyrolyzed photoresist films (PPF) are introduced as planar carbon electrodes in a PDMS−quartz hybrid microchip device. The utility of PPF in electroanalytical applications is demonstrated by the separation and detection of various neurotransmitters. PPF is found to form a stable, low-capacitance, durable layer on quartz, which can then be used in conjunction with a microchip capillary electrophoretic device. Sinusoidal voltammetric detection at PPF electrodes is shown to be very sensitive, with a detection limit (S/N = 3) of 100 nM for dopamine, corresponding to a mass detection limit (S/N = 3) of 2 amol. The selectivity of analysis in the frequency domain is demonstrated by isolating each individual signal in a pair of analytes that are chromatographically unresolved. Effectively decoupling the electrophoresis and electrochemical systems allows the electrodes to be placed just inside the separation channel, which results in efficient separations (80 000−100 000 plates/m).Keywords
This publication has 24 references indexed in Scilit:
- Reliable glucose monitoring through the use of microsystem technologyAnalytical and Bioanalytical Chemistry, 2002
- Sinusoidal voltammetry:Journal of Electroanalytical Chemistry, 2002
- A Chip-Based Electrophoresis System with Electrochemical Detection and Hydrodynamic InjectionAnalytical Chemistry, 2002
- Fully Integrated On-Chip Electrochemical Detection for Capillary Electrophoresis in a Microfabricated DeviceAnalytical Chemistry, 2002
- Capillary Gel Electrophoresis with Sinusoidal Voltammetric Detection: A Strategy To Allow Four-“Color” DNA SequencingAnalytical Chemistry, 2001
- Photoresist-Derived Carbon for Microelectromechanical Systems and Electrochemical ApplicationsJournal of the Electrochemical Society, 2000
- Rapid Prototyping of Microfluidic Systems in Poly(dimethylsiloxane)Analytical Chemistry, 1998
- Optimization of confocal epifluorescence microscopy for microchip-based miniaturized total analysis systemsThe Analyst, 1998
- Counting Single Chromophore Molecules for Ultrasensitive Analysis and Separations on Microchip DevicesAnalytical Chemistry, 1998
- Scanning Tunneling Microscopy of Ordered Graphite and Glassy Carbon Surfaces: Electronic Control of Quinone AdsorptionLangmuir, 1994