Abstract
Background Diffuse reflectance spectroscopy (DRS) uses the steady‐state diffuse reflectance measured from the tissue surface to determine absorption and scattering properties of sampled tissue. Many inverse models used to determine absorber properties have assumed a homogeneous distribution of blood. However, blood in tissue is confined to blood vessels that occupy a small fraction of the overall volume. This simplified assumption can lead to large errors when measuring optical properties. The objective of this study was to examine the effect of confining absorbers to small volumes, such as the microvasculature, on in vivo DRS. Study Design We fabricated multi‐layer microfluidic devices to mimic blood vessels with a size similar to skin microvasculature. We studied the effect of varying channel size (diameter = 22 and 44 µm) and absorber concentration (10–80% food color dye in water) on diffuse reflectance measurements. We also examined the in vivo reflectance from normal skin and non‐melanoma skin cancer on 14 patients. Results Our results demonstrate that both absorption coefficient and vessel diameter affect the diffuse reflectance spectra. An empirically calculated packaging correction factor based on our experiments shows good agreement with previous theoretical derivations of the same factor. In vivo measurements on normal skin and basal cell carcinoma show that incorporating a correction factor greatly improves the fit of the inverse model to the spectra. In addition, there were statistically significant differences in measured mean vessel diameter and blood volume fraction between normal skin and basal cell carcinoma. Conclusion We have demonstrated experimentally the effect of pigment packaging in blood vessels over a physiologically relevant range of blood vessel size and absorption. The correction factors implemented to account for the packaging effect could potentially be used as diagnostic parameters for diagnosing skin cancers. Lasers Surg. Med. 42:680–688, 2010.Keywords
This publication has 19 references indexed in Scilit:
- Re-evaluation of model-based light-scattering spectroscopy for tissue spectroscopyJournal of Biomedical Optics, 2009
- Analysis of changes in reflectance measurements on biological tissues subjected to different probe pressuresJournal of Biomedical Optics, 2008
- Lookup table–based inverse model for determining optical properties of turbid mediaJournal of Biomedical Optics, 2008
- Fabrication of microfluidic systems in poly(dimethylsiloxane)Electrophoresis, 1999
- Predictions and measurements of scattering and absorption over broad wavelength ranges in tissue phantomsApplied Optics, 1997
- Modelling light distributions of homogeneous versus discrete absorbers in light irradiated turbid mediaPhysics in Medicine & Biology, 1997
- Therapeutic response during pulsed laser treatment of port-wine stains: Dependence on vessel diameter and depth in dermisLasers in Medical Science, 1995
- Influence of blood vessels on the measurement of hemoglobin oxygenation as determined by time-resolved reflectance spectroscopyMedical Physics, 1995
- A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivoMedical Physics, 1992
- The flattering of the absorption spectrum of suspensions, as compared to that of solutionsBiochimica et Biophysica Acta, 1956