Experimental analysis of fluid mechanical energy losses in aortic valve stenosis: Importance of pressure recovery
Authors:
- 1 November 1996
- journal article
- Published by Springer Nature in Annals of Biomedical Engineering
- Vol. 24 (6) , 685-694
Abstract
Current methods for assessing the severity of aortic stenosis depend primarily on measures of maximum systolic pressure drop at the aortic valve orifice and related calculations such as valve area. It is becoming increasingly obvious, however, that the impact of the obstruction on the left ventricle is equally important in assessing its severity and could potentially be influenced by geometric factors of the valve, causing variable degrees of downstream pressure recovery. The goal of this study was to develop a method for measuring fluid mechanical energy losses in aortic stenosis that could then be directly related to the hemodynamic load placed on the left ventricle. A control volume form of conservation of energy was theoretically analyzed and modified for application to aortic valve stenosis measurements.In vitro physiological pulsatile flow experiments were conducted with different types of aortic stenosis models, including a venturi meter, a nozzle, and 21-mm Medtronic-Hall tilting disc and St. Jude bileaflet mechanical valves. The energy loss created by each model was measured for a wide range of experimental conditions, simulating physiological variation. In all cases, there was more energy lost for the nozzle (mean=0.27 J) than for any other model for a given stroke volume. The two prosthetic valves generated approximately the same energy losses (mean=0.18 J), which were not statistically different, whereas the venturi meter had the lowest energy loss for all conditions (mean=0.037 J). Energy loss correlated poorly with orifice pressure drop (r 2=0.34) but correlated well with recovered pressure drop (r 2=0.94). However, when the valves were considered separately, orifice and recovered pressure drop were both strongly correlated with energy loss (r 2=0.99, 0.96). The results show that recovered pressure drop, not orfice pressure drop, is directly related to the energy loss that determines pump work and therefore is a more accurate measure of the hemodynamic significance of aortic stenosis.Keywords
This publication has 33 references indexed in Scilit:
- Is Cardiac Catheterization Necessary in the Era of Echocardiography?Echocardiography, 1995
- Pressure recovery in aortic stenosis: An in vitro study in a pulsatile flow modelJACC, 1992
- Aortic valve resistance as an adjunct to the gorlin formula in assessing the severity of aortic stenosis in symptomatic patientsJACC, 1992
- Hemodynamic progression of aortic stenosis in adults assessed by doppler echocardiographyJACC, 1989
- Theoretical evaluation of energy loss methods in the analysis of prosthetic heart valvesJournal of Biomedical Engineering, 1987
- Pressure drops across prosthetic aortic heart valves under steady and pulsatile flow—In vitro measurementsJournal of Biomechanics, 1979
- The Role of Simulation Studies in Cardiac Valve Prosthesis DesignEngineering in Medicine, 1978
- Percentage of Left Ventricular Stroke Work LossCirculation, 1967
- Principles of Fluid Mechanics Applied to Some Situations in the Human Circulation and particularly to the Testing of Valves in a Pulse DuplicatorThorax, 1964
- Hydraulic formula for calculation of the area of the stenotic mitral valve, other cardiac valves, and central circulatory shunts. IAmerican Heart Journal, 1951