Marked discrepancy in pressure gradient between Doppler and catheter examinations on Medtronic Mosaic valve in aortic position
Authors:
- 12 June 2012
- journal article
- case report
- Published by Springer Nature in General Thoracic and Cardiovascular Surgery
- Vol. 60 (12) , 818-821
Abstract
A 71-year-old woman underwent aortic valve replacement with 23 mm Medtronic Mosaic Ultra valve 4 years ago because of aortic stenosis. Although she had been asymptomatic since the operation, echocardiography showed 4 m/s of transprosthetic valve flow that implied early prosthetic valve failure. Catheter examination revealed that the mean transvalvular pressure gradient during systole was 15.1 mmHg on simultaneous pressure recording, and calculated valve area 1.82 cm2. Her body surface area was 1.56 m2. Prosthetic valve failure and prosthesis-patient mismatch were both denied. The discrepancy between Doppler study data and catheter data seemed to be due to fluid dynamical pressure recovery phenomenon. Net pressure difference between the left ventricle and the aorta may be significantly smaller than that estimated using Bernoulli’s equation from transvalvular flow speed in some patients after aortic valve replacement.Keywords
This publication has 6 references indexed in Scilit:
- Impact of Prosthesis–Patient Size on Functional Recovery After Aortic Valve ReplacementCirculation, 2005
- Net pressure gradients in aortic prosthetic valves can be estimated by DopplerJournal of the American Society of Echocardiography, 2003
- “overestimation” of catheter gradients by doppler ultrasound in patients with aortic stenosis: a predictable manifestation of pressure recoveryJACC, 1999
- Importance of Pressure Recovery for the Assessment of Aortic Stenosis by Doppler UltrasoundCirculation, 1996
- Discrepancies between Doppler and catheter gradients in aortic prosthetic valves in vitro. A manifestation of localized gradients and pressure recovery.Circulation, 1990
- The problem of valve prosthesis-patient mismatch.Circulation, 1978