Corneal Biomechanical Changes After Intracorneal Ring Segment Implantation in Keratoconus
Authors:
- Jorge L. AlioJAJorge L. Alio
- Rafael I. BarraquerRBRafael I. Barraquer
- Ralph MichaelRMRalph Michael
- 1 May 2012
- journal article
- Published by Wolters Kluwer Health in Cornea
- Vol. 31 (5) , 491-499
Abstract
To evaluate by means of the Ocular Response Analyzer (ORA) the biomechanical changes that follow intracorneal ring segment (ICRS) implantation and to develop a predicting model for the postoperative visual outcome, considering these biomechanical changes and other clinical parameters. A total of 45 consecutive keratoconic eyes of 35 patients aged 18 to 55 years and implanted with ICRS were included in this retrospective study. All patients were implanted with KeraRing using femtosecond laser technology. Visual acuity, refraction, corneal topography, and aberrations were evaluated during a 6-month follow-up. Additionally, corneal biomechanical changes were evaluated with the ORA system. No significant changes in corneal hysteresis (CH) and corneal resistance factor (CRF) were observed at 1 month postoperatively (P ≥ 0.39). However, significant changes in these parameters were detected afterward (CH, 3 months, P = 0.03; CRF, 6 months, P = 0.02). Preoperative corneal biomechanical parameters were significantly correlated with postoperative corneal higher-order aberrations at all visits (P ≤ 0.05), although these correlations became stronger at the end of the follow-up. Multiple regression analysis revealed that corrected distance visual acuity at 1 month after surgery was significantly correlated with the preoperative mean keratometry and the preoperative difference between CH and CRF (P < 0.01, adjusted R² of 0.66). Biomechanical parameters measured with the ORA and mean keratometry are factors allowing a prediction of the postoperative visual outcome in a short term after ICRS implantation in keratoconic eyes. In the long term, corneal biomechanical changes limit the prediction of the ring segment effect.Keywords
This publication has 39 references indexed in Scilit:
- Biomechanical and morphological corneal response to placement of intrastromal corneal ring segments for keratoconusJournal of Cataract & Refractive Surgery, 2009
- Pentacam posterior and anterior corneal aberrations in normal and keratoconic eyesClinical and Experimental Optometry, 2009
- Rapid visual recovery after penetrating keratoplasty for keratoconusClinical & Experimental Ophthalmology, 2008
- Intracorneal ring segment implantation for the management of keratoconus: Safety and efficacyJournal of Cataract & Refractive Surgery, 2007
- Corneal biomechanical properties in normal, post-laser in situ keratomileusis, and keratoconic eyesJournal of Cataract & Refractive Surgery, 2007
- Assessment of the Biomechanical Properties of the Cornea with the Ocular Response Analyzer in Normal and Keratoconic EyesInvestigative Ophthalmology & Visual Science, 2007
- Biomechanics and wound healing in the corneaExperimental Eye Research, 2006
- Intracorneal ring segments for keratoconus correction: Long-term follow-upJournal of Cataract & Refractive Surgery, 2006
- Analysis of results related to good and bad outcomes of Intacs implantation for keratoconus correctionJournal of Cataract & Refractive Surgery, 2006
- European clinical evaluation: Use of Intacs for the treatment of keratoconusJournal of Cataract & Refractive Surgery, 2006