Optimization of deterministic transport parameters for the calculation of the dose distribution around a high dose‐rate brachytherapy source
Authors:
- Kent A. GiffordKGKent A. Gifford
- Michael J. PriceMPMichael J. Price
- John L. HortonJHJohn L. Horton
- Todd A. WareingTWTodd A. Wareing
- Firas MourtadaFMFiras Mourtada
- 12 May 2008
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 35 (6Part1) , 2279-2285
Abstract
The goal of this work was to calculate the dose distribution around a high dose‐rate brachytherapy source using a multi‐group discrete ordinates code and then to compare the results with a Monte Carlo calculated dose distribution. The unstructured tetrahedral mesh discrete ordinates code Attila™ version 6.1.1 was used to calculate the photon kerma rate distribution in water around the Nucletron microSelectron mHDRv2 source. MCNPX 2.5.c was used to compute the Monte Carlo water photon kerma rate distribution. Two hundred million histories were simulated, resulting in standard errors of the mean of less than overall. The number of energy groups, (angular order), (scattering order), and mesh elements were varied in addition to the method of analytic ray tracing to assess their effects on the deterministic solution. Water photon kerma rate matrices were exported from both codes into an in‐house data analysis software. This software quantified the percent dose difference distribution, the number of points within and , and the mean percent difference between the two codes. The data demonstrated that a 5 energy‐group cross‐section set calculated results to within of a 15 group cross‐section set. was sufficient to resolve the solution in angle. expansion of the scattering cross‐section was necessary to compute accurate distributions. A computational mesh with 55 064 tetrahedral elements in a 30 cm diameter phantom resolved the solution spatially. An efficiency factor of 110 with the above parameters was realized in comparison to MC methods. The Attila™ code provided an accurate and efficient solution of the Boltzmann transport equation for the mHDRv2 source.Keywords
This publication has 16 references indexed in Scilit:
- Comparison of a finite-element multigroup discrete-ordinates code with Monte Carlo for radiotherapy calculationsPhysics in Medicine & Biology, 2006
- Dosimetric evaluation of the Fletcher–Williamson ovoid for pulsed-dose-rate brachytherapy: a Monte Carlo studyPhysics in Medicine & Biology, 2005
- Distortions induced by radioactive seeds into interstitial brachytherapy dose distributionsMedical Physics, 2004
- Reconstruction of electron spectra using singular component decompositionMedical Physics, 2002
- Multigroup discrete ordinates modeling of 125I 6702 seed dose distributions using a broad energy-group cross section representationMedical Physics, 2002
- Fast iterative methods for discrete-ordinates particle transport calculationsProgress in Nuclear Energy, 2002
- Dosimetric modeling of the microselectron high-dose rate 192Ir source by the multigroup discrete ordinates methodMedical Physics, 2000
- Monte Carlo‐aided dosimetry of a new high dose‐rate brachytherapy sourceMedical Physics, 1998
- Dose calculations about shielded gynecological colpostatsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1990