Active Middle Ear Implant Application in Case of Stapes Fixation
Authors:
- 1 September 2010
- journal article
- otosclerosis
- Published by Wolters Kluwer Health in Otology & Neurotology
- Vol. 31 (7) , 1027-1034
Abstract
Hypothesis: Driving the oval window directly with an active middle ear implant (AMEI) can produce high levels of input to the inner ear. Background: Treatment of otosclerosis bypasses the stapes with a piston that penetrates the vestibule. Although this treats the conductive component of hearing loss, it does not treat the sensorineural part, which can be improved using an additional conventional hearing aid. Active middle ear implants have been proposed to be an alternative in treating otosclerosis in cases of mixed hearing losses. Methods: Seven temporal bones were prepared to expose the stapes and round window (RW). Stapes and RW velocities were measured while driving with an AMEI the stapes head with a bell-shaped tip. The stapes footplate was then fixed with acrylic cement; fixation was confirmed through attenuated RW velocities. A cylinder tip (0.5 mm) was then used to drive the inner ear through a stapedotomy with and without interposition of fascia. Results: Driving the stapes with an AMEI produced mean maximum equivalent ear canal sound pressure levels (SPL) of 138 dB (0.25-8 kHz at 1 V [RMS]). Stapes fixation caused a ∼25-dB attenuation. Driving with a cylinder tip through the stapedotomy produced 114 dB SPL (24 dB less than normal) and 110 dB SPL (28 dB less than normal) performance with and without fascia, respectively. Performance with fascia was greater than without. Conclusion: Driving the oval window with an AMEI in a scenario of stapes fixation was demonstrated to be feasible, with performance comparable to traditional AMEI coupling to the incus or stapes. These possibilities offer new perspectives to treat mixed hearing loss in case of fixed footplate.Keywords
This publication has 55 references indexed in Scilit:
- Performance considerations of prosthetic actuators for round-window stimulationHearing Research, 2010
- Evaluation of Round Window Stimulation Using the Floating Mass Transducer by Intracochlear Sound Pressure Measurements in Human Temporal BonesOtology & Neurotology, 2010
- Round Window Membrane Implantation with an Active Middle Ear Implant: A Study of the Effects on the Performance of Round Window Exposure and Transducer Tip Diameter in Human Cadaveric Temporal BonesAudiology and Neurotology, 2010
- Audiological Application Criteria for Implantable Hearing Aid Devices: A Clinical Experience at the Nijmegen ORL ClinicThe Laryngoscope, 2008
- Tolerable Hearing Aid Delays. V. Estimation of Limits for Open Canal FittingsEar & Hearing, 2008
- Conductive Hearing Loss Caused by Third-Window Lesions of the Inner EarOtology & Neurotology, 2008
- Testing a Method for Quantifying the Output of Implantable Middle Ear Hearing DevicesAudiology and Neurotology, 2007
- Evaluation of implant losses and skin reactions around extraoral bone‐anchored implants: A 0‐ to 8‐year follow‐upOtolaryngology–Head and Neck Surgery, 2000
- A Dynamic and Harmonic Damped Finite Element Analysis Model of StapedotomyActa Oto-Laryngologica, 1997
- Reported benefits and problems experienced with bone-anchored hearing aidsBritish Journal of Audiology, 1996