Accuracy of the Withings BPM Connect Device for Self-Blood Pressure Measurements in General Population – Validation According to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard
Open Access
- 1 March 2022
- journal article
- research article
- Published by Taylor & Francis in Vascular Health and Risk Management
- Vol. ume 18, 191-200
Abstract
Background: As recommended by various authorities and scientific societies, electronic devices for blood pressure (BP) measurements must undergo independent clinical validations for accuracy assessment. Objective: To assess the accuracy of the Withings BPM Connect device in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization “AAMI/ESH/ISO Universal Standard (ISO 81060– 2:2018)”. Methods: The Withings BPM Connect device measures BP at the brachial level using the oscillometric method. According to this protocol using the same-arm sequential BP measurement method, subjects (n ≥ 85) fulfilling the age, gender, BP and cuff distribution criteria must be included. Two criteria are used for the analysis: Criterion 1: differences between observers’ mercury sphygmomanometer reference measurements and test device BP values (test versus reference) and their standard deviation (SD) must be ≤ 5 ± 8 mmHg for both systolic (SBP) and diastolic BP (DBP). Criterion 2: the SD of the mean BP differences between the test device and reference BP per subject must be ≤ 6.91/6.62 mmHg for SBP/DBP. Results: Ninety-two subjects were selected, 85 of whom were included. For validation criterion 1, the mean difference ± SD between the reference and device BP values was 0.6 ± 5.3 mmHg for SBP and 2.1 ± 4.3 mmHg for DBP. For criterion 2, the SD of the mean BP differences between the test device and reference BP per subject was 4.2/3.6 mmHg (SBP/DBP). These results fulfilled the AAMI/ESH/ISO Universal Standard (ISO 81060– 2:2018) protocol requirements. Conclusion: The Withings BPM Connect oscillometric device for home BP measurement fulfilled all of the accuracy requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060– 2:2018) in general population and consequently can be recommended for home BP measurements.Keywords
This publication has 12 references indexed in Scilit:
- Home blood pressure monitoring: methodology, clinical relevance and practical application: a 2021 position paper by the Working Group on Blood Pressure Monitoring and Cardiovascular Variability of the European Society of HypertensionJournal Of Hypertension, 2021
- 2021 European Society of Hypertension practice guidelines for office and out-of-office blood pressure measurementJournal Of Hypertension, 2021
- STRIDE BP: an international initiative for accurate blood pressure measurementJournal Of Hypertension, 2020
- Lancet Commission on Hypertension group position statement on the global improvement of accuracy standards for devices that measure blood pressureJournal Of Hypertension, 2020
- Recommendations and Practical Guidance for performing and reporting validation studies according to the Universal Standard for the validation of blood pressure measuring devices by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO)Journal Of Hypertension, 2019
- 2018 ESC/ESH Guidelines for the management of arterial hypertensionJournal Of Hypertension, 2018
- A universal standard for the validation of blood pressure measuring devicesJournal Of Hypertension, 2018
- A Universal Standard for the Validation of Blood Pressure Measuring DevicesHypertension, 2018
- European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adultsBlood Pressure Monitoring, 2010
- Short report: An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devicesJournal Of Hypertension, 1993