Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers
Authors:
- 20 April 2011
- journal article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 49 (7) , 1187-1191
Abstract
Background: The random use of point-of-care, blood gas and core laboratory analyzers to measure electrolytes and metabolites increases the variability in test results. This study was designed to determine whether these results performed on whole blood (point-of-care and blood gas) and plasma (core laboratory) platforms are interchangeable without a risk of clinically relevant discrepancies. Methods: The interchangeability of the blood gas analysis, electrolytes, glucose, lactate and hemoglobin results performed with three stat platforms (i-STAT, Radiometer ABL 825, RapidLab 865) and two core laboratory platforms (Roche Modular P800 and Sysmex XE-2100) were evaluated using samples from critically ill patients. Results: For pH, pCO2, pO2 and Ca2+, good correlation (r-values 0.96–1.00) was observed for all comparative analyzers and the biases were within clinically acceptable limits. Potassium, sodium, glucose, lactate and hemoglobin measured with stat analyzers was highly correlated with measurements performed using the laboratory analyzers, r-values 0.89–1.00 and slopes 0.83–1.07. Mean differences with significant bias (pConclusions: The blood gas, K+, Na+, Ca2+, glucose and lactate results measured with stat and core laboratory analyzers can be used in different clinical settings for critical care management. However, when monitoring small changes in sodium concentrations, the use of single analyzer is encouraged to avoid analytical differences (acceptance limit ±2%) and misinterpretation of results measured with multiple analyzers. The users of the i-STAT at low hemoglobin values overdiagnose anaemia. Thus, prior to transfusion, the use of hemoglobin concentrations measured with laboratory analyzers is preferable.Keywords
This publication has 12 references indexed in Scilit:
- Laboratory testing during critical care transport: point-of-care testing in air ambulancesClinical Chemistry and Laboratory Medicine (CCLM), 2010
- Relationship of the absolute difference between direct and indirect ion selective electrode measurement of serum sodium and the total protein concentrationJournal of Clinical Pathology, 2008
- The Effect of Albumin Concentration on Plasma Sodium and Chloride Measurements in Critically Ill PatientsAnesthesia & Analgesia, 2007
- Sodium measurement: effects of differing sampling and analytical methodsAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2006
- Sodium removal and sodium concentration during peritoneal dialysis: effects of three methods of sodium measurementNephrology Dialysis Transplantation, 2004
- Comparison of Point-of-Care Versus Central Laboratory Measurement of Electrolyte Concentrations on Calculations of the Anion Gap and the Strong Ion DifferenceAnesthesiology, 2003
- Accuracy of the i-STATTM bedside blood gas analyserEuropean Respiratory Journal, 1999
- Quality specifications based on analysis of effects of performance on clinical decision-makingScandinavian Journal of Clinical and Laboratory Investigation, 1999
- ErratumClinical Biochemistry, 1997
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986