Automatic switching between the AAI and the DDD algorithm can prevent repetitive non‐reentrant ventriculoatrial synchrony
Authors:
- Takeshi KitamuraTKTakeshi Kitamura
- Seiji FukamizuSFSeiji Fukamizu
- Masahiro NauchiMNMasahiro Nauchi
- Takuro NishimuraTNTakuro Nishimura
- Tomohiko WatanabeTWTomohiko Watanabe
- Jin IwasawaJIJin Iwasawa
- Hiroshi ShimadaHSHiroshi Shimada
- Tae IshikawaTITae Ishikawa
- Noriko MatsushitaNMNoriko Matsushita
- Tomomi AbeTATomomi Abe
- Rintaro HojoRHRintaro Hojo
- Takekuni HayashiTHTakekuni Hayashi
Open Access
- 29 July 2013
- journal article
- Published by Wiley in Journal of Arrhythmia
- Vol. 30 (2) , 115-118
Abstract
A 67‐year‐old man with non‐obstructive hypertrophic cardiomyopathy had received an implantable cardioverter‐defibrillator (ICD) for an unstable, sustained ventricular tachycardia (VT) induced by programmed stimulation during an electrophysiological study 5 years earlier. An intracardiac electrogram recorded by the ICD revealed repetitive, non‐reentrant ventriculoatrial synchrony (RNRVAS) associated with hypotension. Electrophysiologic and hemodynamic studies indicated that RNRVAS was induced and reproducibly termed by a single ventricular extrastimulus from the right ventricular apex. Following attainment of the elective replacement indicator, we replaced the ICD with another having managed ventricular pacing, which automatically switched AAI and DDD, thereby avoiding unnecessary ventricular pacing. Thus far, the patient has not experienced further RNRVAS. Thus, we believe that automatic switching between AAI and DDD can prevent RNRVAS.Keywords
This publication has 7 references indexed in Scilit:
- Repetitive Non‐reentrant VA Synchrony and Pacemaker‐Mediated Tachycardia Induced by the AF Suppression AlgorithmPacing and Clinical Electrophysiology, 2009
- Percent ventricular pacing with managed ventricular pacing mode in standard pacemaker populationEP Europace, 2008
- Percent right ventricular pacing predicts outcomes in the DAVID trialHeart Rhythm, 2005
- Multicenter, Prospective, Randomized Safety and Efficacy Study of a New Atrial‐Based Managed Ventricular Pacing Mode (MVP) in Dual Chamber ICDsJournal of Cardiovascular Electrophysiology, 2005
- Pacemaker Repetitive Nonreentrant Ventriculoatrial Synchronous Rhythm. A ReviewJournal of Interventional Cardiac Electrophysiology, 2001
- Pacemaker Induced Repetitive Nonreentrant Ventriculoatrial Synchrony: Initiation and Termination by Ventricular ExtrasystolePacing and Clinical Electrophysiology, 1997
- Repetitive reentrant and non‐reentrant ventriculoatrial synchrony in dual chamber pacingClinical Cardiology, 1991