Comparison of third generation versus fourth generation electronic apex locators in detecting apical constriction: An in vivo study

Authors:

Abstract
Aim: The aim of this in vivo study was to compare the accuracy of Root ZX and Raypex 5 in detecting minor diameter in human permanent single-rooted teeth. Materials and Methods: Thirty-one patients with completely formed single-rooted permanent teeth indicated for extraction were selected for the study. Crown was flattened for stable reference point and access cavity prepared. Working length was determined with both apex locators. A 15 K file adjusted to that reading was placed in the root canal and stabilized with cement. The tooth was then extracted atraumatically. Following extraction apical 4 mm of root was shaved. The position of the minor diameter in relation to the anatomic apex was recorded for each tooth under stereomicroscope at ×10. The efficiency of two electronic apex locators to determine the minor diameter was statistically analyzed using paired sample t-test. Results: The minor diameter was located within the limits of ±0.5 mm in 96.6% of the samples with the Root ZX and 93.2% of the samples with Raypex 5. The paired sample t-test showed no significant difference. Conclusion: On analyzing the results of our study it can be concluded that Raypex 5 was as effective as Root ZX in determining the minor diameter.

This publication has 17 references indexed in Scilit: