Thermal voltage converters for accurate voltage measurements to 30 megacycles per second

Authors:

Abstract
Thermal voltage converters, each consisting of a resistor in series with a thermoelement in a coaxial line, have been developed for measurements of rms voltages of 1 to 200 volts at frequencies from 3 cps (cycles per second) to 30 mc (megacycles). An accuracy of 0.1% or better may be obtained by a-c-d-c transfer techniques up to at least 10 mc and 0.2% at 30 mc.