Abstract
We have demonstrated a novel Fourier-domain optical coherence tomography system and signal-processing algorithm for full-range, real-time, artifact-free quantitative imaging of the anterior chamber. Cross-sectional full-range images comprising 1024 × 800 pixels (axial × lateral) were acquired and displayed at 6.7 images/s. Volumetric data comprising 1024 × 400 × 60 pixels (axial × lateral × elevation) were acquired in 4.5 seconds with real-time visualization of individual slices and 3-dimensional reconstruction performed in postprocessing. Details of the cornea, limbus, iris, anterior lens capsule, trabecular meshwork, and Schlemm's canal were visualized. Quantitative surface height maps of the corneal epithelium and endothelium were obtained from the volumetric data and used to generate corneal thickness maps.Keywords
This publication has 37 references indexed in Scilit:
- Fourier Domain Mode Locking (FDML): A new laser operating regime and applications for optical coherence tomographyOptics Express, 2006
- Visual Optics: The Shapes of PupilsCurrent Biology, 2006
- Improved complex spectral domain OCT for in vivo eye imagingOptics Communications, 2005
- Swept source optical coherence tomography using an all-fiber 1300-nm ring laser sourceJournal of Biomedical Optics, 2005
- Heterodyne swept-source optical coherence tomography for complete complex conjugate ambiguity removalJournal of Biomedical Optics, 2005
- Ultrahigh-resolution high-speed retinal imaging using spectral-domain optical coherence tomographyOptics Express, 2004
- Ultrahigh-resolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensationOptics Express, 2004
- Instantaneous quadrature low-coherence interferometry with 3×3 fiber-optic couplersOptics Letters, 2003
- Spectral resolution and sampling issues in Fourier-transform spectral interferometryJournal of the Optical Society of America B, 2000
- Measurement of intraocular distances by backscattering spectral interferometryOptics Communications, 1995